Recall Z-2269-2025
Beckman Coulter, Inc. · initiated June 25, 2025
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
UniCel DxI 600 Access Immunoassay Analyzer, Part Number A30260
Codes / lots: UDI-DI: 15099590341602; Serial Numbers: 902467 902468 902465 902464 902466 902469 902476 902477 902479 902480 902481 902482 902483 902485 902484 902486 902487 902489 902313 902310 902384 902536 902347 902584 902355 902394 902541 902054 902058 902324 902090 902403 902417 902471 902517 902563 902080 902565 902320 902308 902388 902440 902330 902337 902503 902557 902279 902287 902269 902513 902507 902543 902449 902494 902300 902286 902250 902259 902562 902506 902508 902311 902560 902207 902587 902345 902574 902208 902211 902243 902262 902263 902295 902533 902572 902470 902319 902336 902251 901723 902344 902195 902282 902280 902448 902458 902381 902406 902372 902198 902540 902199 902234 902209 902217 902276 902569 902568 902577 902510 902078 902438 902056 902203 902205 902593 902439 902380 902397 902309 902196 902410 902402 902367 902585 902241 902423 902343 902595 902206 902210 902338 902202 902442 902445 902074 902076 902359 902288 902073 902072 902400 902580 902389 902567 902573 902461 902396 902570 902571 902221 902538 902411 902473 902495 902496 902200 902201 902472 902349 902084 902258 902341 902497 902553 902552 902299 902499 902512 902498 902215 902267 902266 902255 902214…
Reason for recall
Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed.
FDA-determined cause: Device Design
Action taken
An URGENT MEDICAL DEVICE RECALL notification dated 6/25/25 was sent to customers. ACTION: If your laboratory encounters any of the events described in this letter: " Perform a system initialization to resume normal operation. o If repeated initialization attempts fail, contact Customer Technical Support. " Repeat testing for any cancelled tests. " No action is required if your laboratory has not observed the events listed above ESOLUTION: "A Beckman Coulter representative will schedule the modification installation when it is available. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product listed above to another laboratory, please provide them a copy of this letter. So that we are assured you have received this important communication, please respond within 10 days in one of the following ways: " Electronically, if you received this communication via email. " Manually, complete and return the enclosed Response Form. If you have any questions regarding this notice, please contact our Customer Support: " From our website: http://www.beckmancoulter.com
Details
- Recall number
- Z-2269-2025
- Initiated
- June 25, 2025
- Posted by FDA
- August 5, 2025
- Product code
- JJE – Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Quantity
- 271 systems
- Distribution
- Worldwide distribution - US Nationwide.
- Recalling firm
- Beckman Coulter, Inc., Chaska, MN
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-2266-2025 Class II | UniCel DxI 800 Access Immunoassay Analyzer, Part Number 973100Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running… | 06/25/2025Open, Classified |
| Z-2267-2025 Class II | DxI 800 Access Immunoassay Analyzer W/Dual Gantry, Part Number A71457Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running… | 06/25/2025Open, Classified |
| Z-2268-2025 Class II | DxI 800 Access Immunoassay Analyzer W/Spot B, Part Number A71456Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running… | 06/25/2025Open, Classified |
| Z-2270-2025 Class II | DxI 600 Access Immunoassay Analyzer W/Dual Gantry, Part Number A71461Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running… | 06/25/2025Open, Classified |
| Z-2271-2025 Class II | DxI 600 Access Immunoassay Analyzer W/Spot B, Part Number A71460Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running… | 06/25/2025Open, Classified |