Recall Z-2273-2016

Ortho-Clinical Diagnostics · initiated May 18, 2016

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

VITROS Chemistry Products Ca Slides, GEN 29, REF/Product Code 145 0261, Unique Device Identifier No. 10758750009114 or 20758750009111, IVD -- Sales Unit: 300 slides/pack Product Usage: For in vitro diagnostic use only. VITROS Ca Slides quantitatively measure calcium (Ca) concentration in serum, plasma, and urine. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone disorders, chronic renal disease and tetany (intermittent muscular contractions or spasms).

Codes / lots: Lots #s: 0329-0528-8193 (exp. 05/01/2017), 0329-0528-8892 (exp. 06/01/2017), 0329-0528-8192 (exp. 05/01/2017), 0329-0528-8902 (exp. 06/01/2017), 0329-0528-8226 (exp. 05/01/2017), 0329-0528-8928 (exp. 07/01/2017), 0329-0528-8227 (exp. 05/01/2017), 0329-0528-9000 (exp. 07/01/2017), 0329-0528-8359 (exp. 05/01/2017), 0329-0528-8927 (exp. 07/01/2017), 0329-0528-8361 (exp. 05/01/2017), 0329-0528-8965 (exp. 07/01/2017), 0329-0528-8360 (exp. 05/01/2017), 0329-0528-8966 (exp. 07/01/2017), 0329-0528-8324 (exp. 05/01/2017), 0329-0528-9167 (exp. 07/01/2017), 0329-0528-8323 (exp. 05/01/2017), 0329-0528-9171 (exp. 07/01/2017), 0329-0528-8325 (exp. 05/01/2017), 0329-0528-9196 (exp. 07/01/2017), 0329-0528-6999 (exp. 03/01/2017), 0329-0528-9172 (exp. 07/01/2017), 0329-0528-7000 (exp. 03/01/2017), 0329-0528-9177 (exp. 07/01/2017), 0329-0528-6998 (exp. 03/01/2017), 0329-0528-9265 (exp. 07/01/2017), 0329-0528-8903 (exp. 06/01/2017), 0329-0528-9176 (exp. 07/01/2017), 0329-0528-8901 (exp. 06/01/2017), 0329-0528-9264 (exp. 07/01/2017), 0329-0528-8893 (exp. 06/01/2017), 0329-0528-9368 (exp. 07/01/2017), 0329-0528-8891 (exp. 06/01/2017), 0329-0528-9369 (exp. 07/01/2017).

Reason for recall

Ortho Clinical Diagnostics (OCD) confirmed a potential for biased calcium (Ca) results to be generated when using VITROS Ca Slides, GEN29, and the VITROS 250/350/950/5,1 FS, 4600 Chemistry System or the VITROS 5600 Integrated System. An increase in negatively biased calcium results complaints prompted the investigation that identified this product issue.

FDA-determined cause: Nonconforming Material/Component

Action taken

Ortho Clinical Diagnostics sent an URGENT PRODUCT CORRECTION NOTIFICATION letter dated May 18, 2016 to affected customers. . The letter identified the affected product, problem and actions to be taken. The letter instructed to discontinue use of the affected lot(s) and discard all remaining inventory. Customers were asked to complete and return the Confirmation of Receipt form. For questions contact Customer Technical Services at 1-800-421-3311.

Details

Recall number
Z-2273-2016
Initiated
May 18, 2016
Terminated
June 8, 2018
Product code
CJY – Azo Dye, Calcium
Quantity
US: 27,625 sales units; Foreign: 4757 sales units
Distribution
Worldwide Distribution - US Nationwide in the countries of Puerto Rico and Foreign distribution to the following countries: Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, Poland and Venezuela.
Recalling firm
Ortho-Clinical Diagnostics, Rochester, NY