Recall Z-2277-2023

GE OEC Medical Systems, Inc · initiated May 15, 2023

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

OEC 9900 Systems with 9-inch Image Intensifier

Codes / lots: UDI-DI: 00840682114301 Serial Numbers: ES-2700, ESXXXL00081, ESXXXL00104, ES-2873, ESXXXX04525, ESXXXX04526, ESXXXX04527, E9-2573, E9XXXX03024, ESXXXX04467, ES-2768, ESXXXX03911, ESXXXX04425, E9-2510, E9-2646, ES-3208, ES-3209, ES-2934, ES-2940, ESXXXX04391, ES-0662, ES-3216, ES-3217, E9-1266, E9-2330, ESXXXX04441, E9XXXX03364, ES-1631, ES-2835, ES-2846, ES-2987, ES-3179, ESXXXX04377, ES-2692, ES-2693, ES-2847, ES-2679, E9-1695, E9-1716, E9-1719, E9-2183, E9-2186, E9-2327, E9-2328, E9-2451, E9-2517, ES-3110, ESXXXX04365, ESXXXX04369, E9-2286, E9-2571, ESXXXX04472, ES-2647, ESXXXX04436, E9-2189, ES-2755, ESXXXX04475, E9-2645, ESXXXX04449, ES-2667, ESXXXX04435, E9XXXX03408, E9-2187, ES-2868, ESXXXX04394, ES-2981, ES-2942, ES-2965, ESXXXX04505, E9-2246, ES-2988, E9XXXX03336, ES-1548, ES-2948, ESXXXX04462, E9-0975, ES-3080, ESXXXX04061, ESXXXX04497, ESXXXX04465, ESXXXX04468, ESXXXX04489, E9-2562, ES-2953, ESXXXX04464, ESXXXX04488, ESXXXX04484, ES-2808, ES-2689, ES-2848, ES-2995, ESXXXL00057, ES-2849, ES-3090, ES-2836, E9XXXX03484, ESXXXX04447, ES-2978, ESXXXL00083, ESXXXX04530, ES-2640, ES-2865, ESXXXX04457, ES-2894, ES-3091, ES-3076, ES-2788, ESXXXX04460, ESXXXX04406, E9-2282…

Reason for recall

Image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which could lead to injuries such as fractures, hematoma, contusion, bruise, pain.

FDA-determined cause: Process control

Action taken

On 5/15/23 correction notices were mailed to: Director/Manager of Radiology, Hospital Administrator, Head of Radiology Department, PACS Administrator, Director of IT Department, Head, Biomedical Engineering, who were asked to do the following: You can continue to use your device. When transporting the device, ensure it is always in the Anterior Posterior (AP) position, as instructed in the systems Operator Manual. Use caution when moving system to avoid a large impact force. In addition, customers were asked to complete and return the acknowledgement response form to fieldactionssurgery@ge.com Customers with questions or concerns can contact the firm at 1-800-437-1171

Details

Recall number
Z-2277-2023
Initiated
May 15, 2023
Posted by FDA
July 31, 2023
Product code
OXO – Image-Intensified Fluoroscopic X-Ray System, Mobile
Quantity
908
Distribution
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and the countries of Algeria, Argentina, Australia, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria…
Recalling firm
GE OEC Medical Systems, Inc, Salt Lake City, UT

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2275-2023 Class IIOEC Flexiview 8800 Systems with 9-inch Image IntensifierImage intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the… 05/15/2023Open, Classified
Z-2276-2023 Class IIOEC 9800 Systems with 9-inch Image IntensifierImage intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the… 05/15/2023Open, Classified
Z-2278-2023 Class IIOEC Elite Systems with 9-inch Image IntensifierImage intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the… 05/15/2023Open, Classified
Z-2279-2023 Class IIField Upgrade Kits: P/N 5451208 RIU, 9 INCH, TOSHIBA; P/N 5485304 SERVICE PART, IMAGE…Image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the… 05/15/2023Open, Classified