Recall Z-2278-2025

Boston Scientific Corporation · initiated June 25, 2025

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

FARAWAVE 1.0 Pulsed Field Ablation Catheter, Product ID M004PF41M401

Codes / lots: UDI-DI: 00191506043148; Batch Numbers: 36543581 36543644 36543646 36568553 36572565 36572568 36572571 36572576 36578308 36578314 36579110 36598352 36598809 36599402 36600406 36600543 36600544 36600545 36600578

Reason for recall

The potentially impacted units were manufactured using specific equipment that may have caused cracks in the electrode bands on the catheter's distal end.

FDA-determined cause: Nonconforming Material/Component

Action taken

An Urgent Medical Device Removal - Immediate Action Required notification letter dated June 2025 was sent to customers. Actions 1. Do NOT use affected product (Attachment 1), remove affected devices from your facility s inventory, segregate the units in a secure place until they can be returned to Boston Scientific. 2. Immediately post this information in a visible location near the affected products to ensure this information is readily accessible to all handlers and users of the device. 3. Forward this notice to any healthcare professional from your organization for awareness and to any organization where affected devices have been transferred. 4. Complete and return the enclosed Reply Verification Tracking Form per the enclosed instructions on page five. 5. Return affected product. Email: BSCFieldActionCenter@bsci.com Reach out to your local Boston Scientific representative with any questions.

Details

Recall number
Z-2278-2025
Initiated
June 25, 2025
Posted by FDA
August 8, 2025
Product code
QZI – Percutaneous Cardiac Ablation Catheter For Treatment Of Atrial Fibrillation With Irreversible Electroporation
Quantity
445 units
Distribution
US Nationwide.
Recalling firm
Boston Scientific Corporation, Saint Paul, MN

Official FDA recall record