Recall Z-2287-2008

Abaxis Inc · initiated April 21, 2008

Terminated

Product

Piccolo Chemistry Analyzer software version 6.121 with Electrolyte Panel Lot #8041AC5, model number 400-1022 in polyfoil pouch. Piccolo Electrolyte Panel is a single-use, self-contained reagent disk intended for use with the Piccolo and PiccoloXpress chemistry analyzers, for in vitro qualitative determination.

Codes / lots: software version 6.121 with Electrolyte Panel Lot #8041AC5, expiration date 2009-01-21

Reason for recall

The product will recover tCO2 at approximately 5-7 units high relative to some controls and patient samples

FDA-determined cause: Process control

Action taken

The firm issued a notification letter to its consignees on April 21, 2008. The firm intends to follow up with phone calls and provide a software upgrade for the products. Contact Abaxis Customer Service at 1-510-675-6500 for assistance.

Details

Recall number
Z-2287-2008
Initiated
April 21, 2008
Posted by FDA
September 17, 2008
Terminated
April 26, 2011
Product code
JLB – Atp And Creatine Kinase (Enzymatic), Creatine
Quantity
237 kits
Distribution
Product was distributed Worldwide.
Recalling firm
Abaxis Inc, Union City, CA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2286-2008 Piccolo Chemistry Analyzer software version 6.121 with Comprehensive Metabolic Lot…The product will recover tCO2 at approximately 5-7 units high relative to some controls and patient samples 04/21/2008Terminated