Recall Z-2296-2008

Zimmer Inc. · initiated July 7, 2008

Terminated

Product

Zimmer Patient Helper Reinforced Overhead Bar for Hill-Rom beds with removable headboards; Zimmer Inc., Warsaw, IN; Prod. Nos. 00-0965-000-00 and 00-0965-001-00. The product is a L shaped bar that is used with hospital beds set up for traction. The bar allows mounting of a trapeze to help the patient pull themselves up when in the hospital bed.

Codes / lots: All units shipped between 7/11/07 and 5/19/08.

Reason for recall

If the bed is in the upright position and the bar is dropped rapidly or in free-fall, it could strike the patient in the bed, and the drop could present a laceration hazard by creating a sharp edge on the metal bar.

FDA-determined cause: Device Design

Action taken

Consignees were notified via recall letter dated 7/7/08 to discontinue use of these devices until the firm visits and replaces the devices. Contact Zimmer at 1-800-321-5533 for assistance.

Details

Recall number
Z-2296-2008
Initiated
July 7, 2008
Posted by FDA
September 16, 2008
Terminated
November 12, 2009
Product code
IKX – Aid, Transfer
Quantity
645 of all systems
Distribution
Nationwide and Canada.
Recalling firm
Zimmer Inc., Warsaw, IN

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2297-2008 Zimmer Patient Helper Reinforce Overhead Bar for Hill-Rom Versa Care beds; Zimmer Inc.…If the bed is in the upright position and the bar is dropped rapidly or in free-fall, it could strike the… 07/07/2008Terminated
Z-2298-2008 Zimmer Freestanding Patient Helper for Hill-Rom Totalcare beds; Zimmer Inc., Warsaw, IN…If the bed is in the upright position and the bar is dropped rapidly or in free-fall, it could strike the… 07/07/2008Terminated