Recall Z-2296-2024
American Contract Systems Inc · initiated May 15, 2024
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
TOTAL KNEE - 297835-Procedure Kit Catalog Number: RGTK10K
Codes / lots: UDI-DI: 00191072203267 LOT#"s: 665231
Reason for recall
Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.
FDA-determined cause: Under Investigation by firm
Action taken
American Contract Systems issued Urgent Voluntary Recall Letter (FA-2024-018) via email on 5/15/24 to Distributors. Distributors to notify end-users. Letter states reason for recall, health risk and action to take: Immediately discontinue use of the cast padding component (DISCARD ITEM 4927009046-CAST PADDING, 6" Cotton Blend Rayon) included in each kit/tray identified in ATTACHMENT 1. Please review your inventory, and segregate and quarantine all affected Trays. If any of the affected trays identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. "Add warning labels to all kits/cases impacted: A labeling template is enclosed for printing warning labels which should be affixed to all affected product in inventory. This label should be affixed to each affected kit and case in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing labeling. "Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. "Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. "Complete the enclosed Response Form and return as soon as possible. "If credit is requested, please send an email to quality@owens-minor.com.
Details
- Recall number
- Z-2296-2024
- Initiated
- May 15, 2024
- Posted by FDA
- July 5, 2024
- Product code
- OJH – Orthopedic Tray
- Quantity
- 24 units
- Distribution
- US Nationwide distribution.
- Recalling firm
- American Contract Systems Inc, Tiffin, OH
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-2274-2024 Class II | AHAR50K STANDARD ARTHROSCOPY PACK - 291852. Procedure tray Catalog Number: AHAR50KEthylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual… | 05/15/2024Open, Classified |
| Z-2275-2024 Class II | TOTAL KNEE - 302497- Procedure tray Catalog Number: ANTK12AFEthylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual… | 05/15/2024Open, Classified |
| Z-2276-2024 Class II | MAJOR KNEE - 302497- Procedure tray Catalog Number: ASKN23BEthylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual… | 05/15/2024Open, Classified |
| Z-2277-2024 Class II | LOWER EXTREMITY - 247542- Procedure tray Catalog Number: BFLE56XEthylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual… | 05/15/2024Open, Classified |
| Z-2278-2024 Class II | TOTAL KNEE - 247546 - Procedure tray Catalog Number: BFTK01AAEthylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual… | 05/15/2024Open, Classified |
| Z-2279-2024 Class II | TOTAL KNEE - 247546 Procedure tray Catalog Number: BFTK01ZEthylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual… | 05/15/2024Open, Classified |
| Z-2280-2024 Class II | ARTHROSCOPY WOLFSON PACK BHS Procedure tray Catalog Number: BPAR65AEthylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual… | 05/15/2024Open, Classified |
| Z-2281-2024 Class II | ARTHROSCOPY SUP JOI PACK Procedure tray Catalog Number: BPAS17CEthylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual… | 05/15/2024Open, Classified |
| Z-2282-2024 Class II | ARTHO KNEE BAPTIST SOUTH PACK Procedure tray Catalog Number: BPKA49BEthylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual… | 05/15/2024Open, Classified |
| Z-2283-2024 Class II | ARTHRO KNEE BEACHES PACK BHS- Procedure tray Catalog Number: BPKB46BEthylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual… | 05/15/2024Open, Classified |
| Z-2284-2024 Class II | BPKN15C KNEE BASIN PACK BHS- Procedure tray Catalog Number: BPKN15CEthylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual… | 05/15/2024Open, Classified |
| Z-2285-2024 Class II | FOOT ANKLE-LF Procedure tray Catalog Number:CCFA51BEthylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual… | 05/15/2024Open, Classified |
| Z-2286-2024 Class II | HAND WRIST FOREARM-LF Procedure tray Catalog Number: CCHD17BEthylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual… | 05/15/2024Open, Classified |
| Z-2287-2024 Class II | KNEE SCOPE Procedure tray Catalog Number: COKN48LEthylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual… | 05/15/2024Open, Classified |
| Z-2288-2024 Class II | UPPER EXTREMITY PK, STRL F G Catalog Number: IHUE40AGEthylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual… | 05/15/2024Open, Classified |
| Z-2289-2024 Class II | ACL, ST LUKES METHODIST 1CS Catalog Number: LMAC10AAEthylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual… | 05/15/2024Open, Classified |
| Z-2290-2024 Class II | TOTAL KNEE, ST LUKES Catalog Number: LMTK17APEthylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual… | 05/15/2024Open, Classified |
| Z-2291-2024 Class II | KNEE ARTHROSCOPY PACK Catalog Number: MEKA12REthylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual… | 05/15/2024Open, Classified |
| Z-2292-2024 Class II | KNEE ARTHROSCOPY-Procedure Kit Catalog Number: QPKA35AEthylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual… | 05/15/2024Open, Classified |
| Z-2293-2024 Class II | TOTAL KNEE -Procedure Kit Catalog Number: QPTK95AEthylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual… | 05/15/2024Open, Classified |
| Z-2294-2024 Class II | TOTAL KNEE -Procedure Kit Catalog Number: QPTK95BEthylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual… | 05/15/2024Open, Classified |
| Z-2295-2024 Class II | HAND PACK 297829--Procedure Kit Catalog Number: RGHN04IEthylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual… | 05/15/2024Open, Classified |
| Z-2297-2024 Class II | TOTAL KNEE - 297835 -Procedure Kit Catalog Number: RGTK10LEthylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual… | 05/15/2024Open, Classified |
| Z-2298-2024 Class II | KNEE ARTHROSCOPY PACK-Procedure Kit Catalog Number: SFAR60AEthylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual… | 05/15/2024Open, Classified |
| Z-2299-2024 Class II | TOTAL KNEE-Procedure Kit Catalog Number: SFTK54BEthylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual… | 05/15/2024Open, Classified |
| Z-2300-2024 Class II | HAND DRAPING SJH-Procedure Kit Catalog Number: SJHD67FEthylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual… | 05/15/2024Open, Classified |
| Z-2301-2024 Class II | KNEE ARTHROSCOPY-Procedure Kit Catalog Number: SLAR14GEthylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual… | 05/15/2024Open, Classified |
| Z-2302-2024 Class II | TOTAL KNEE-Procedure Kit Catalog Number: SMTK21CEthylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual… | 05/15/2024Open, Classified |
| Z-2303-2024 Class II | ARTHROSCOPY PACK-Procedure Kit Catalog Number: TNAR22UEthylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual… | 05/15/2024Open, Classified |
| Z-2304-2024 Class II | ARTHROSCOPY PACK-Procedure Kit Catalog Number: TNAR22VEthylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual… | 05/15/2024Open, Classified |