Recall Z-2299-2009

Ebi, L.P. · initiated April 2, 2009

Terminated

Product

Bone Screw4.5MM X 95MM Bone Screw Lock, Rx only Biomet, 100 Interpace Parkway, Parsippany, NJ 070454 The screw devices are an internal fixation device intended to aid in the alignment and stabilization of fractures to the skeletal system until healing has occurred. Implants used in this application include non sterile and sterile bone screws.

Codes / lots: Item number 27895, lot number 717050

Reason for recall

Some devices could have been made with raw material lots that may have inclusions, and could result in decreased strength of the device.

FDA-determined cause: Nonconforming Material/Component

Action taken

The firm sent Urgent Recall Notices to all direct accounts, hospital Risk Management and Physicians by first class mail. Letters are dated March 29, 2009, April 2, 2009 and April 16, 2009 respectively. The letter describes the problem and that that there is a possible risk of a device failure. An additional e-mail was sent on May 22, 2009 to 5 customers due to an expansion of the recall to include 4 additional items. On June 19, 2009, the firm again expanded the recall to include 2 additional products and notified customers. Questions should be directed to William Hennig, at 1.973.299.9300. ext. 1557.

Details

Recall number
Z-2299-2009
Initiated
April 2, 2009
Posted by FDA
September 22, 2009
Terminated
September 23, 2009
Product code
HSB – Rod, Fixation, Intramedullary And Accessories
Quantity
2
Distribution
Worldwide - US -NY, NJ, MI, MD, FL, GA, PR, MN, SD, OK, TX,, LA, CA, PA, OH, NV, HI, ME, NC, KS, UT, CO, TN, WV, VA, and WA. Foreign: Brazil, Colombia, Denmark, Germany, Finland, Greece, Italy, Japan, The Netherlands, Switzerland, Sweden, Portugal, Spain, United Kingdom, Saudi Arabia, and Trinidad.
Recalling firm
Ebi, L.P., Parsippany, NJ

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2277-2009 TI Cann Partl Thd Recon Screw 6.00MM x 65MM; Trauma Fixation Systems, Rx Only; Biomet…Some devices could have been made with raw material lots that may have inclusions, and could result in… 04/02/2009Terminated
Z-2278-2009 TI Cann Partl Thd Recon Screw 6.00MM x 70MM; Trauma Fixation Systems, Rx Only; Biomet…Some devices could have been made with raw material lots that may have inclusions, and could result in… 04/02/2009Terminated
Z-2279-2009 TI Cann Partl Thd Recon Screw 6.00MM x 75MM; Trauma Fixation Systems, Rx Only; Biomet…Some devices could have been made with raw material lots that may have inclusions, and could result in… 04/02/2009Terminated
Z-2280-2009 TI Cann Partl Thd Recon Screw 6.00MM x 80MM; Trauma Fixation Systems, Rx Only; Biomet…Some devices could have been made with raw material lots that may have inclusions, and could result in… 04/02/2009Terminated
Z-2281-2009 TI Cann Partl Thd Recon Screw 6.00MM x 85MM; Trauma Fixation Systems, Rx Only; Biomet…Some devices could have been made with raw material lots that may have inclusions, and could result in… 04/02/2009Terminated
Z-2282-2009 TI Cann Partl Thd Recon Screw 6.00MM x 90MM; Trauma Fixation Systems, Rx Only; Biomet…Some devices could have been made with raw material lots that may have inclusions, and could result in… 04/02/2009Terminated
Z-2283-2009 TI Cann Partl Thd Recon Screw 6.00MM x 95MM; Trauma Fixation Systems, Rx Only; Biomet…Some devices could have been made with raw material lots that may have inclusions, and could result in… 04/02/2009Terminated
Z-2284-2009 TI Cannulated Screw 20MM thrd 6.5MM (d) X 85MM (L); Trauma Fixation Systems, Rx Only…Some devices could have been made with raw material lots that may have inclusions, and could result in… 04/02/2009Terminated
Z-2285-2009 TI Cannulated Screw 32MM thrd 7.3MM (d) X 95MM (L); Trauma Fixation Systems, Rx Only…Some devices could have been made with raw material lots that may have inclusions, and could result in… 04/02/2009Terminated
Z-2286-2009 TI Cannulated Screw 32MM thrd 7.3MM (d) X 100MM (L); Trauma Fixation Systems, Rx Only…Some devices could have been made with raw material lots that may have inclusions, and could result in… 04/02/2009Terminated
Z-2287-2009 TI Cannulated Screw 32MM thrd 7.3MM (d) X 120MM (L); Trauma Fixation Systems, Rx Only…Some devices could have been made with raw material lots that may have inclusions, and could result in… 04/02/2009Terminated
Z-2288-2009 TI Screw Buttress Thread 5MM (D) x 35MM (L) Trauma Fixation Systems, Rx Only; Biomet…Some devices could have been made with raw material lots that may have inclusions, and could result in… 04/02/2009Terminated
Z-2289-2009 TI Screw Buttress Thread 6MM (D) x 65MM (L) Trauma Fixation Systems, Rx Only; Biomet…Some devices could have been made with raw material lots that may have inclusions, and could result in… 04/02/2009Terminated
Z-2290-2009 Bone Screw 5.5MM X 45MM Bone Screw Lck-Cann, Rx only Biomet, 100 Interpace Parkway…Some devices could have been made with raw material lots that may have inclusions, and could result in… 04/02/2009Terminated
Z-2291-2009 Bone Screw 5.5MM X 55MM Bone Screw Lck-Cann, Rx only Biomet, 100 Interpace Parkway…Some devices could have been made with raw material lots that may have inclusions, and could result in… 04/02/2009Terminated
Z-2292-2009 Bone Screw 5.5MM X 70MM Bone Screw Lck-Cann, Rx only Biomet, 100 Interpace Parkway…Some devices could have been made with raw material lots that may have inclusions, and could result in… 04/02/2009Terminated
Z-2293-2009 Bone Screw 4.5MM X 20MM Bone Screw Lock, Rx only Biomet, 100 Interpace Parkway…Some devices could have been made with raw material lots that may have inclusions, and could result in… 04/02/2009Terminated
Z-2294-2009 Bone Screw 4.5MM X 28MM Bone Screw Lock, Rx only Biomet, 100 Interpace Parkway…Some devices could have been made with raw material lots that may have inclusions, and could result in… 04/02/2009Terminated
Z-2295-2009 Bone Screw 4.5MM X 36MM Bone Screw Lock, Rx only Biomet, 100 Interpace Parkway…Some devices could have been made with raw material lots that may have inclusions, and could result in… 04/02/2009Terminated
Z-2296-2009 Bone Screw 4.5MM X 52MM Bone Screw Lck-Cann, Rx only Biomet, 100 Interpace Parkway…Some devices could have been made with raw material lots that may have inclusions, and could result in… 04/02/2009Terminated
Z-2297-2009 Bone Screw .5MM X 60MM Bone Screw Lock, Rx only Biomet, 100 Interpace Parkway…Some devices could have been made with raw material lots that may have inclusions, and could result in… 04/02/2009Terminated
Z-2298-2009 Bone Screw 4.5MM X 70MM Bone Screw Lock, Rx only Biomet, 100 Interpace Parkway…Some devices could have been made with raw material lots that may have inclusions, and could result in… 04/02/2009Terminated
Z-2300-2009 Cannulated Screw Ballista Percutaneous Pedicle Screw System. 6/5 x 45MM; Biomet Spine…Some devices could have been made with raw material lots that may have inclusions, and could result in… 04/02/2009Terminated
Z-2301-2009 C-Tek Maxan Anterior Cervical Plate System; 4.0MM X 20MM Fixed Screw' Biomet Spine, 100…Some devices could have been made with raw material lots that may have inclusions, and could result in… 04/02/2009Terminated
Z-2302-2009 Pediatric Femoral Nail 5.5MM (d) x 28 CM (L) Rx only, Sterile; Biomet, 100 Interpace…Some devices could have been made with raw material lots that may have inclusions, and could result in… 04/02/2009Terminated
Z-2303-2009 Pediatric Femoral Nail 5.5MM (d) x 30 CM (L) Rx only, Sterile; Biomet, 100 Interpace…Some devices could have been made with raw material lots that may have inclusions, and could result in… 04/02/2009Terminated
Z-2304-2009 Pediatric Femoral Nail 5.5MM (d) x 32 CM (L) Rx only, Sterile; Biomet, 100 Interpace…Some devices could have been made with raw material lots that may have inclusions, and could result in… 04/02/2009Terminated
Z-2305-2009 Pediatric Femoral Nail 5.5MM (d) x 36 CM (L) Rx only, Sterile; Biomet, 100 Interpace…Some devices could have been made with raw material lots that may have inclusions, and could result in… 04/02/2009Terminated
Z-2306-2009 Fixed Closed Screw, 6.5mm (D) x 40mm (L), Made in USA; Biomet Spine, Parsippany, NJ The…Some devices could have been made with raw material lots that may have inclusions, and could result in… 04/02/2009Terminated
Z-2307-2009 Fixed Open Screw (TI); 7.0MM (D) x 30MM (L); Made In USA, Biomet Spine, Parsippany, NJ…Some devices could have been made with raw material lots that may have inclusions, and could result in… 04/02/2009Terminated