Recall Z-2312-2010

Physio Control, Inc. · initiated November 17, 2008

Terminated

Product

Pediatric Paddles attachments used with LIFEPAK 9, LIFEPAK 9P, LIFEPAK 12 and LIFEPAK 20/20e Defibrillator/Monitors.

Codes / lots: The product does not have serial numbers or lot numbers. The product has the Part Number 800418-XX (XX symbolize any 2 numbers).

Reason for recall

Sterilization efficacy could not be confirmed for sterilization methods on External Sterilizable Paddles and Pediatric Paddle attachments.

FDA-determined cause: Labeling False and Misleading

Action taken

On 11/14/08, Physio Control began sending the Urgent User Information letter via certified mail to their customers. The letter notified the customers about the modification to the cleaning and sterilization instruction of the External Sterilization Defibrillation Paddles and Pediatric Paddle attachments used with LIFEPAK 9, LIFEPAK 9P, LIFEPAK 12, and LIFEPAK 20/20e Defibrillator/Monitors The customers are informed to discontinue use of the gas sterilization methods listed as 12/88 Ethylene Oxide (EO) or 100% Ethylene Oxide (OE). They are notified that they can continue to use the sterilization method such as STERRAD per existing Operating Instructions, but limit sterilization cycles from 100 to 50 cycles. The customers are advised that prior to and after each use, they should examine cables and connector for damage or signs of wear such as loose connections, damaged pins, exposed wires and cable connector corrosion and examine paddles for scratched or pitted surfaces and stop using product if damage has occurred. Within 45 days, the firm will update the Operating Instruction that will include revised Sterilization Guidelines on a website www.physio-control.com/products/product-support. Customers can call Technical Support at 1-800-442-1142, #5 or they can visit the firm's website at www.physio-control-notices.com/extpaddles regarding the recall.

Details

Recall number
Z-2312-2010
Initiated
November 17, 2008
Posted by FDA
August 28, 2010
Terminated
September 28, 2010
Product code
LDD – Dc-Defibrillator, Low-Energy, (Including Paddles)
Quantity
1689 units
Distribution
Devices were distributed to medical facilities through out the US and WORLD WIDE to Canada, Germany, India, Netherlands, Australia, Hong Kong, Japan, Brazil, Panama, Russia, Korea, Malaysia, Venezuela, and United Arab Emirates(UAE) .
Recalling firm
Physio Control, Inc., Redmond, WA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2311-2010 External Sterilizable Paddles for use with the LIFEPAK 9, LIFEPAK 9P, LIFEPAK 12, and…Sterilization efficacy could not be confirmed for sterilization methods on External Sterilizable Paddles and… 11/17/2008Terminated