Recall Z-2335-2012

Steris Corporation · initiated August 1, 2012

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

The SYSTEM 1E Liquid Sterilant Processing System. The SYSTEM 1E Liquid Sterilant Processing System is intended for the liquid chemical sterilization of manually cleaned immersible and reusable critical and semi-critical heat sensitive medical devices, including endoscopes and their accessories.

Codes / lots: Model #: P6500, S/N 400000 through 406294

Reason for recall

The firm initiated a recall after complaint investigations revealed that when the units are left on and not in use at night and/or on weekends the hose disconnects/separates resulting in water leakage that can cause damage to property.

FDA-determined cause: Nonconforming Material/Component

Action taken

STERIS sent a Urgent Voluntary Field Correction Notice letter dated August 1, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. STERIS has received no reports of injuries associated with the disconnection of a SYSTEM IE water hose. STERIS Corporation will instill new hoses and connections on your SYSTEM IE Liquid Chemical Sterilant Processing System. For further information or if you have any questions regarding our visit to your facility, please contact STERIS Field Service Dispatch at 1-800-333-8828.

Details

Recall number
Z-2335-2012
Initiated
August 1, 2012
Posted by FDA
September 7, 2012
Terminated
June 6, 2014
Product code
MED – Sterilant, Medical Devices
Quantity
6,051 units
Distribution
Worldwide Distribution -- USA (nationwide) and the countries of Bahrain and Hong Kong.
Recalling firm
Steris Corporation, Mentor, OH

Official FDA recall record