Recall Z-2338-2025

Olympus Corporation of the Americas · initiated July 25, 2025

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Electrosurgical, Cutting and Coagulation and Accessories. Electrosurgical Generator ESG-410 Surgical Energy Platform (SEP). Model Number: WA91327U. The ESG-410 Electrosurgical Generator is an electrosurgical generator which utilizes monopolar and bipolar high frequency current and supports ultrasonic instruments during open surgery, laparoscopic surgery, including single-site surgery, and endoscopic surgery.

Codes / lots: Model Number: WA91327U. Lot Code: ESG-410 (SEP) WA91327U. UDI-DI: 04042761087698. Serial Numbers: 101696, 101697, 101698, 101699, 101700, 101701, 101702, 101703, 101704, 101705, 101706, 101707, 101708, 101709, 101710, 101919, 101920, 101921, 101922, 101923, 101924, 101925, 101926, 101927, 101928, 101929, 101933, 101934, 101935, 101936, 101937, 101940, 101941, 101942, 101945, 101946, 101947, 102148, 102149, 102150, 102151, 102152, 102153, 102154, 102156, 102157, 102158, 102159, 102160, 102161, 102162, 102164, 102165, 102167, 102168, 102170, 102172, 102173, 102174, 102175, 102176, 102177, 102178, 102179, 102180, 102181, 102182, 102183, 102184, 102185, 102186, 102187, 102188, 102189, 102190, 102191, 102193, 102194, 102195, 102198, 102199, 102200, 102266, 102427, 101932, 102155, 102163, 102197, 102660, 102661, 102662, 102663, 102664, 102665, 102666, 102667, 102668, 102669, 102670, 102671, 102672, 102673, 102674, 102675, 102676, 102677, 102678, 102679, 102680, 102681, 102682, 102683, 102684, 102685, 102686, 102687, 102688, 102689, 102690, 102691, 102692, 102693, 102694, 102695, 102696, 102697, 102698, 102699, 102700, 102701, 102702, 102703, 102704, 102705, 102706, 102707, 102708…

Reason for recall

Potential for the ESG-410 Electrosurgical Generator to display an error message: "E0662 System Error", followed by either automatic single reboots of the device, or continuous reboot 'loops'.

FDA-determined cause: Component change control

Action taken

Olympus notified consignees via letter on about 07/25/2025. Consignees were instructed to examine inventory, identify any affected units, and continued use is permitted until the repair is performed as long as users follow the IFU. Olympus will contact consignees to arrange for the return of affected units to replace the affected part. Consignees were also instructed to notify customers, if the units were further distributed, and to acknowledge receipt of the letter through the Olympus web portal.

Details

Recall number
Z-2338-2025
Initiated
July 25, 2025
Posted by FDA
August 19, 2025
Product code
GEI – Electrosurgical, Cutting & Coagulation & Accessories
Quantity
155 units
Distribution
US Nationwide distribution.
Recalling firm
Olympus Corporation of the Americas, Center Valley, PA