Recall Z-2338-2025
Olympus Corporation of the Americas · initiated July 25, 2025
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Electrosurgical, Cutting and Coagulation and Accessories. Electrosurgical Generator ESG-410 Surgical Energy Platform (SEP). Model Number: WA91327U. The ESG-410 Electrosurgical Generator is an electrosurgical generator which utilizes monopolar and bipolar high frequency current and supports ultrasonic instruments during open surgery, laparoscopic surgery, including single-site surgery, and endoscopic surgery.
Codes / lots: Model Number: WA91327U. Lot Code: ESG-410 (SEP) WA91327U. UDI-DI: 04042761087698. Serial Numbers: 101696, 101697, 101698, 101699, 101700, 101701, 101702, 101703, 101704, 101705, 101706, 101707, 101708, 101709, 101710, 101919, 101920, 101921, 101922, 101923, 101924, 101925, 101926, 101927, 101928, 101929, 101933, 101934, 101935, 101936, 101937, 101940, 101941, 101942, 101945, 101946, 101947, 102148, 102149, 102150, 102151, 102152, 102153, 102154, 102156, 102157, 102158, 102159, 102160, 102161, 102162, 102164, 102165, 102167, 102168, 102170, 102172, 102173, 102174, 102175, 102176, 102177, 102178, 102179, 102180, 102181, 102182, 102183, 102184, 102185, 102186, 102187, 102188, 102189, 102190, 102191, 102193, 102194, 102195, 102198, 102199, 102200, 102266, 102427, 101932, 102155, 102163, 102197, 102660, 102661, 102662, 102663, 102664, 102665, 102666, 102667, 102668, 102669, 102670, 102671, 102672, 102673, 102674, 102675, 102676, 102677, 102678, 102679, 102680, 102681, 102682, 102683, 102684, 102685, 102686, 102687, 102688, 102689, 102690, 102691, 102692, 102693, 102694, 102695, 102696, 102697, 102698, 102699, 102700, 102701, 102702, 102703, 102704, 102705, 102706, 102707, 102708…
Reason for recall
Potential for the ESG-410 Electrosurgical Generator to display an error message: "E0662 System Error", followed by either automatic single reboots of the device, or continuous reboot 'loops'.
FDA-determined cause: Component change control
Action taken
Olympus notified consignees via letter on about 07/25/2025. Consignees were instructed to examine inventory, identify any affected units, and continued use is permitted until the repair is performed as long as users follow the IFU. Olympus will contact consignees to arrange for the return of affected units to replace the affected part. Consignees were also instructed to notify customers, if the units were further distributed, and to acknowledge receipt of the letter through the Olympus web portal.
Details
- Recall number
- Z-2338-2025
- Initiated
- July 25, 2025
- Posted by FDA
- August 19, 2025
- Product code
- GEI – Electrosurgical, Cutting & Coagulation & Accessories
- Quantity
- 155 units
- Distribution
- US Nationwide distribution.
- Recalling firm
- Olympus Corporation of the Americas, Center Valley, PA
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-2339-2025 Class II | Electrosurgical, Cutting and Coagulation and Accessories. Electrosurgical Generator…Potential for the ESG-410 Electrosurgical Generator to display an error message: "E0662 System Error"… | 07/25/2025Open, Classified |