Recall Z-2352-2010

Abbott Laboratories · initiated April 16, 2010

Terminated

Product

CELL-DYN Ruby, Hematology Analyzer manufactured by Abbott Laboratories, Santa Clara, CA. Waste Reservoir Assembly Part number 8921174802.

Codes / lots: Reservoir Date Codes Dec 17 2009 to March 17, 2010. Part Number/Serial numbers-- 08H67-01/35286BG-35318BG, 34891BG, 34641BG, 34408BG, 34890BG, 34021BG

Reason for recall

Epoxy holding the two parts of the waste reservoir together fails, potentially resulting in biohazardous fluid leaks.

FDA-determined cause: Nonconforming Material/Component

Action taken

Product Recall letters were sent on 04/16/2010. The letters identified the affected product, explained the reason for recall, the patient impact/user safety, stated the necessary actions, and contact information. Users are to discontinue use until the Waste Chamber(s) have been replaced and follow the instructions in the Hazards section. In addition, appropriate protective equipment needs to be worn along with following their recommended laboratory procedures dealing with spills. Users are to contact their local Abbott Customer Support rep. US customers should be direct their questions to Customer Support.

Details

Recall number
Z-2352-2010
Initiated
April 16, 2010
Posted by FDA
September 2, 2010
Terminated
December 6, 2010
Product code
GKZ – Counter, Differential Cell
Quantity
69 units total
Distribution
Worldwide Distribution -- CA, CT, ID, IN, and MA; and, countries of Australia, Brazil, Canada, China, Colombia, Mexico, Puerto Rico, Singapore, and Uruguay.
Recalling firm
Abbott Laboratories, Santa Clara, CA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2350-2010 CELL-DYN 3200 CS and SL, Hematology Analyzer manufactured by Abbott Laboratories, Santa…Epoxy holding the two parts of the waste reservoir together fails, potentially resulting in biohazardous… 04/16/2010Terminated
Z-2351-2010 CELL-DYN 3700 CS and SL, Hematology Analyzer manufactured by Abbott Laboratories, Santa…Epoxy holding the two parts of the waste reservoir together fails, potentially resulting in biohazardous… 04/16/2010Terminated
Z-2353-2010 CELL-DYN Sapphire, Hematology Analyzer manufactured by Abbott Laboratories, Santa Clara…Epoxy holding the two parts of the waste reservoir together fails, potentially resulting in biohazardous… 04/16/2010Terminated