Recall Z-2354-2018

FHC, Inc. · initiated May 10, 2018

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

microTargeting" Lead Adapter for STar" Drive, Product Number 70-CN-DB

Codes / lots: UDI: 00873263002643

Reason for recall

The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.

FDA-determined cause: Device Design

Action taken

Urgent Medical Device Recall notification letters were sent to customers on 5/10/18. The letters included the following instructions: FHC, Inc. Field Representatives will review your inventory and replace effected Depth Stop Adapters you have on hand with an improved design. In the interim, we advise that you do not over-tighten the Depth Stop Adapter screw. If low impedances or a short circuit is found, physicians should replace the lead prior to use and report the issue to FHC, Inc. or your local field rep.

Details

Recall number
Z-2354-2018
Initiated
May 10, 2018
Terminated
April 30, 2019
Product code
HAW – Neurological Stereotaxic Instrument
Quantity
2952 total
Distribution
The devices were distributed to the following US states: AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The devices were distributed to the following foreign countries: Argentina, Australia, Canada, Columbia, Denmark, Ecuador, France, Germany, Great…
Recalling firm
FHC, Inc., Bowdoin, ME

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2345-2018 Class IImicroTargeting" Drive System, Product Number MT-DSThe FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is… 05/10/2018Terminated
Z-2346-2018 Class IImicroTargeting" Drive System, Product Number MT-DS-01The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is… 05/10/2018Terminated
Z-2347-2018 Class IImicroTargeting" Drive System, Product Number 66-ZD-MDThe FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is… 05/10/2018Terminated
Z-2348-2018 Class IImicroTargeting" Drive System, Product Number 66-ZD-MD-01The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is… 05/10/2018Terminated
Z-2349-2018 Class IImicroTargeting" STar" Drive System (Manual), Product Number ST-DS-MAThe FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is… 05/10/2018Terminated
Z-2350-2018 Class IImicroTargeting" STar" Drive System (Motorized), Product Number ST-DS-METhe FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is… 05/10/2018Terminated
Z-2351-2018 Class IImicroTargeting" STar" Drive System (Manual), Product Number 70-ZD-MAThe FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is… 05/10/2018Terminated
Z-2352-2018 Class IImicroTargeting" STar" Drive System (Motorized), Product Number 70-ZD-METhe FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is… 05/10/2018Terminated
Z-2353-2018 Class IImicroTargeting" Lead Adapter for microTargeting" Drive, Product Number 66-CN-DBThe FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is… 05/10/2018Terminated
Z-2355-2018 Class IImicroTargeting" Drive System Distributed by Medtronic, Product Number FC1006The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is… 05/10/2018Terminated
Z-2356-2018 Class IImicroTargeting" Drive System Distributed by Medtronic, Product Number 9033G0601The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is… 05/10/2018Terminated
Z-2357-2018 Class IImicroTargeting" STar" Drive System (Manual) Distributed by Medtronic, Product Number…The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is… 05/10/2018Terminated
Z-2358-2018 Class IImicroTargeting" STar" Drive System (Motorized) Distributed by Medtronic, Product Number…The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is… 05/10/2018Terminated
Z-2359-2018 Class IImicrotargeting" Depth Stop Adapter, Product Model Number, Product Number 66-AC-DS(1.8),The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is… 05/10/2018Terminated