Recall Z-2365-2024

Abbott Medical · initiated May 16, 2024

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Abbott Proclaim Plus 5 Implantable Pulse Generator, REF 3671, Sterile EO, Rx Only

Codes / lots: 05415067046390/ All Serial Numbers

Reason for recall

The duration between the implantable pulse generator reaching the elective replacement indicator threshold and end of service may be shorter than indicated in product labeling.

FDA-determined cause: Vendor change control

Action taken

On June 6, 2024, Abbott issued a Urgent Medical Device Correction notification to affected consignees. Abbott asked consignees to take the following actions: 1. This issue will continue to affect your patients for the foreseeable future. Abbott is working on an update to align the product labeling with actual device performance. 2. When an ERI notification appears, see the following considerations to make a determination with your patients on timing of IPG replacement within the customer notification. 3. See the customer notification for recommended strategies for preventing lapse in therapy. 4. Please return a completed Acknowledgement Form and maintain a record of this notice along with a copy of the completed Acknowledgement Form. 5. For questions about this issue or reported complaints, please contact your Abbott representative, or Abbott Technical Support at 1-800-727-7846. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax.

Details

Recall number
Z-2365-2024
Initiated
May 16, 2024
Posted by FDA
July 16, 2024
Product code
LGW – Stimulator, Spinal-Cord, Totally Implanted For Pain Relief
Quantity
38 units
Distribution
Worldwide - US Nationwide distribution including in the states of ALABAMA, ALASKA, ARIZONA, ARKANSAS, CALIFORNIA, COLORADO, CONNECTICUT, DELAWARE, DISTRICT OF COLUMBIA, FLORIDA, GEORGIA, GUAM, HAWAII, IDAHO, ILLINOIS, INDIANA, IOWA, KANSAS, KENTUCKY, LOUISIANA,MAINE, MARYLAND, MASSACHUSETTS, MICHIGAN, MINNESOTA, MISSISSIPPI, MISSOURI, MONTANA, NEBRASKA, NEVADA, NEW HAMPSHIRE, NEW JERSEY, NEW…
Recalling firm
Abbott Medical, Plano, TX

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2358-2024 Class IIAbbott Proclaim XR 5 Implantable Pulse Generator, REF 3660, Sterile EO, Rx OnlyThe duration between the implantable pulse generator reaching the elective replacement indicator threshold… 05/16/2024Open, Classified
Z-2359-2024 Class IIAbbott Proclaim 5 Implantable Pulse Generator, REF 3661, Sterile EO, Rx OnlyThe duration between the implantable pulse generator reaching the elective replacement indicator threshold… 05/16/2024Open, Classified
Z-2360-2024 Class IIAbbott Proclaim XR 7 Implantable Pulse Generator, REF 3662, Sterile EO, Rx OnlyThe duration between the implantable pulse generator reaching the elective replacement indicator threshold… 05/16/2024Open, Classified
Z-2361-2024 Class IIAbbott Proclaim 7 Implantable Pulse Generator, REF 3663, Sterile EO, Rx OnlyThe duration between the implantable pulse generator reaching the elective replacement indicator threshold… 05/16/2024Open, Classified
Z-2362-2024 Class IIAbbott Proclaim 5 Implantable Pulse Generator, REF 3665, Sterile EO, Rx OnlyThe duration between the implantable pulse generator reaching the elective replacement indicator threshold… 05/16/2024Open, Classified
Z-2363-2024 Class IIAbbott Proclaim 7 Implantable Pulse Generator, REF 3667, Sterile EO, Rx OnlyThe duration between the implantable pulse generator reaching the elective replacement indicator threshold… 05/16/2024Open, Classified
Z-2364-2024 Class IIAbbott Proclaim Plus 5 Implantable Pulse Generator, REF 3670, Sterile EO, Rx OnlyThe duration between the implantable pulse generator reaching the elective replacement indicator threshold… 05/16/2024Open, Classified
Z-2366-2024 Class IIAbbott Proclaim Plus 7 Implantable Pulse Generator, REF 3672, Sterile EO, Rx OnlyThe duration between the implantable pulse generator reaching the elective replacement indicator threshold… 05/16/2024Open, Classified
Z-2367-2024 Class IIAbbott Proclaim DRG Implantable Pulse Generator, REF 3664, Sterile EO, Rx OnlyThe duration between the implantable pulse generator reaching the elective replacement indicator threshold… 05/16/2024Open, Classified
Z-2368-2024 Class IIProclaim 7 Elite Implantable Pulse Generator, REF 3662, Sterile EO, Rx OnlyThe duration between the implantable pulse generator reaching the elective replacement indicator threshold… 05/16/2024Open, Classified
Z-2369-2024 Class IIProclaim 5 Elite Implantable Pulse Generator, REF 3660, Sterile EO, Rx OnlyThe duration between the implantable pulse generator reaching the elective replacement indicator threshold… 05/16/2024Open, Classified