Recall Z-2368-2020

Resource Optimization & Innovation LLC · initiated May 4, 2020

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Regard SRHS GIC Port A Cath, Sterile, Qty 20/case - Product Usage: The Saline Flush Syringes are intended to be used only for the flushing of indwelling of vascular access devices.

Codes / lots: Item number 303394REG: lot numbers: 259427, 262441, 264440, 268589, 268915, 271008, and 269694.

Reason for recall

There is a potential for the product to exhibit holes in the packaging, which impacts package integrity and potentially compromises a sterile field.

FDA-determined cause: Under Investigation by firm

Action taken

The firm's supplier mailed an Urgent: Field Action Notification dated 04/21/2020 The notification informs the customers to remove the affected product from inventory and place it in quarantine immediately. The notification included a label template for printing the appropriate warning label. The customers will affix the warning label on all affected kits affected by the notification in inventory. The label should be applied to each affected convenience kit and product case in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. Labels can be printed. The customers should send quantity and end user information request to the contact listed. The kits do NOT need to be returned and may be used. The letter states, an alternate saline flush solution should be used to accompany patient procedures until field action inventory is depleted. The notification states, if any of the affected products identified above were further distributed, to ensure all end-users are appropriately notified of the product recall. Recall Response Forms should be completed and returned as soon as possible even if customers have no current product inventory.

Details

Recall number
Z-2368-2020
Initiated
May 4, 2020
Terminated
September 23, 2022
Product code
OFF – Central Venous Catheter Tray
Quantity
9,360 units
Distribution
US Nationwide distribution including in the states of MO, NC.
Recalling firm
Resource Optimization & Innovation LLC, Saint Louis, MO

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2369-2020 Class IIRegard SRHS Central Line Kit, Sterile, 20/case - Product Usage: The Saline Flush…There is a potential for the product to exhibit holes in the packaging, which impacts package integrity and… 05/04/2020Terminated