Recall Z-2376-2021

Smiths Medical ASD Inc. · initiated August 6, 2021

Class I Open, Classified

Class I – dangerous or defective products that could cause serious health problems or death

Product

NORMOFLO Irrigation Fluid Warmer Models H-1100 and H-1129

Codes / lots: Model H-1100: Product Codes CON-H1100, H-1100, H-1100-ES-115V, H-1100-ES-230V, H-1100-FR-230V, H-1100-INT-230, H-1100-IT-230V, H-1100-NL-230V H-1100-SV-230V, H-1100-UK-230 Model H-1129: Product Code H-1129

Reason for recall

There is potential for harmful levels of aluminum leaching when using lower flow rates with certain solutions and blood products and longer duration of use.

FDA-determined cause: Device Design

Action taken

Smiths Medical sent an Urgent Medical Device Recall Notice and attachments dated August 13, 2021 to all affected customers. The letter instructed all customers who purchased affected devices listed in the table on page 1 of the Urgent Medical Device Recall Notice must identify any of the affected products within their possession and refer to the detailed instructions on pages 2 and 3 of the notice. Distributors will be instructed to immediately forward a copy of the Urgent Medical Device Recall Notice and attachments to any customers to whom they have distributed affected product. Distributors will also be instructed to request that customers to who they have distributed affected product complete the Response Form and return it to the distributor. Distributors will indicate their identity as the distributor and document the consignees name and address. Questions regarding this recall notification may be forwarded to fieldactions@smiths-medical.com.

Details

Recall number
Z-2376-2021
Initiated
August 6, 2021
Product code
LGZ – Warmer, Thermal, Infusion Fluid
Quantity
913
Distribution
US Nationwide Distribution and PR OUS distribution in - Austria, Australia, Belgium, Switzerland, Canada, Germany, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Hungary, Ireland, Italy, Kuwait, Luxembourg, Netherlands, New Zealand, Panama, Poland, Portugal, Sweden, Turkey, Venezuela, South Africa
Recalling firm
Smiths Medical ASD Inc., Minneapolis, MN

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2377-2021 Class INORMOFLO Irrigation Warming administration set (disposable), Models IR-40, IR-500…There is potential for harmful levels of aluminum leaching when using lower flow rates with certain… 08/06/2021Open, Classified