Recall Z-2383-2020

Stryker GmbH · initiated May 7, 2020

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Stryker Locking Screw, Fully Threaded T2 Tibia 5x37,5 mm Catalog Number: 18965037S - Product Usage: Stryker implants are single use devices intended for the temporary fixation, correction or stabilization of bones.

Codes / lots: Lot Number: K029C38, K029C39, K030EF6, K030EF7, K030EF9, K031B66

Reason for recall

Seal integrity of the blister pack may be compromised and sterility cannot be assured

FDA-determined cause: Process control

Action taken

Stryker issued Urgent Medical Device Recall (UMDR) - PFA 2365134 letter on May 7, 2020 to affected US consignees by third-party recall company, Stericycle Inc. (2670 Executive Drive Indianapolis, IN 46241), via UPS 2nd day air. An OUS customer letter will also be issued at the country level The letter states reason for recall, heath risk and action to take: 1. Please inform users of this Urgent Medical Device Recall and forward this notice to all individuals who need to be aware within your organization. 2. Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list is at your facility. 3. Quarantine and discontinue use of the recalled devices identified in the affected product list. Return the devices back to Stryker. 4. Under 21 CFR 803, manufactures are also required to report any serious injuries where a product has contributed or may have contributed to the event. Please keep Stryker informed of any adverse events associated with this product by emailing soprodexpreports@stryker.com. 5. A response is required, even though you may not have any physical inventory on site anymore. 6. Hospitals/Branches/Agencies: Complete and sign the enclosed Urgent Medical Device Recall Business Reply Form and fax a copy to 1-888-912-8457 or email to StrykerOrtho8810@Stericycle.com. 7. Hospitals Only: Please contact your Local Sales Office or your Stryker Sales Representative directly for product returns and inventory questions. 8.Branches/Agencies Only: Return all affected devices available at your location to the following address. Stryker Orthopaedics/PFA Product Returns Attn: Distribution Inventory Team 325 Corporate Drive, Dock M-East Mahwah, NJ 07430 Ref. PFA 2365134 Questions or concerns after reviewing the letter, please contact Customer Service at (201) 831-5000.

Details

Recall number
Z-2383-2020
Initiated
May 7, 2020
Posted by FDA
June 16, 2020
Terminated
September 21, 2023
Product code
HSB – Rod, Fixation, Intramedullary And Accessories
Quantity
US 76; OUS 290
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Canada, Chile, China, France, Germany, Italy, Japan, Netherlands, Poland, South Africa, Spain, Sweden, Switzerland, United Kingdom. Updated 7/21/20: Portugal, Denmark, Finland, Norway, Ireland, and Luxembourg.
Recalling firm
Stryker GmbH, Selzach

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2373-2020 Class IIStryker Locking Screw, Fully Threaded S2 05x35 mm Catalog Number: 17965035S - Product…Seal integrity of the blister pack may be compromised and sterility cannot be assured 05/07/2020Terminated
Z-2374-2020 Class IIStryker Compression Screw, Advanced T2 Tibia Catalog Number: 18220001S - Product Usage:…Seal integrity of the blister pack may be compromised and sterility cannot be assured 05/07/2020Terminated
Z-2375-2020 Class IIStryker End Cap T2 Tibia +10 mm Catalog Number: 18220010S - Product Usage: Stryker…Seal integrity of the blister pack may be compromised and sterility cannot be assured 05/07/2020Terminated
Z-2376-2020 Class IIStryker End Cap, Standard T2 Humerus 6 mm Catalog Number: 18300003S - Product Usage:…Seal integrity of the blister pack may be compromised and sterility cannot be assured 05/07/2020Terminated
Z-2377-2020 Class IIStryker 1 End Cap T2 Humerus 6mm, 5mm height Catalog Number: 18300005S - Product Usage:…Seal integrity of the blister pack may be compromised and sterility cannot be assured 05/07/2020Terminated
Z-2378-2020 Class IIStryker End Cap T2 Humerus 6mm, 15mm height Catalog Number: 18300015S - Product Usage:…Seal integrity of the blister pack may be compromised and sterility cannot be assured 05/07/2020Terminated
Z-2379-2020 Class IIStryker Locking Screw, Partially Threaded T2 Tibia 5x55 mm Catalog Number: 18915055S -…Seal integrity of the blister pack may be compromised and sterility cannot be assured 05/07/2020Terminated
Z-2380-2020 Class IIStryker Locking Screw, Fully Threaded T2 Humerus 4x32 mm Catalog Number: 18964032S -…Seal integrity of the blister pack may be compromised and sterility cannot be assured 05/07/2020Terminated
Z-2381-2020 Class IIStryker Locking Screw, Fully Threaded T2 Tibia 5x30 mm Catalog Number: 18965030S -…Seal integrity of the blister pack may be compromised and sterility cannot be assured 05/07/2020Terminated
Z-2382-2020 Class IIStryker Locking Screw, Fully Threaded T2 Tibia 5x35 mm Catalog Number: 18965035S -…Seal integrity of the blister pack may be compromised and sterility cannot be assured 05/07/2020Terminated
Z-2384-2020 Class IIStryker Locking Screw, Fully Threaded T2 Tibia 5x40 mm Catalog Number: 18965040S -…Seal integrity of the blister pack may be compromised and sterility cannot be assured 05/07/2020Terminated
Z-2385-2020 Class IIStryker Locking Screw, Fully Threaded T2 Tibia 5x45 mm Catalog Number: 18965045S -…Seal integrity of the blister pack may be compromised and sterility cannot be assured 05/07/2020Terminated
Z-2386-2020 Class IIStryker Locking Screw, Fully Threaded T2 Tibia 5x50 mm Catalog Number: 18965050S -…Seal integrity of the blister pack may be compromised and sterility cannot be assured 05/07/2020Terminated
Z-2387-2020 Class IIStryker End Cap, Std, Ti Gamma3 Catalog Number: 30051100S - Product Usage: Stryker…Seal integrity of the blister pack may be compromised and sterility cannot be assured 05/07/2020Terminated