Recall Z-2389-2011
Vygon Corporation · initiated April 19, 2011
Product
Churchill Medical Systems AMS-8316CP-1 (Central Line Dressing Kit). The device intended use is a central line dressing kit.
Codes / lots: Lot Numbers: 10I60, 10K42 and 10F87.
Reason for recall
Skin-Prep wipes purchased by Churchill (a Vygon company) for inclusion into Dressing Kits and PICC Insertion Kits were recalled by the manufacturer due to potential bacterial contamination.
FDA-determined cause: Nonconforming Material/Component
Action taken
A recall letter printed on Vygon letterhead dated April 20, 2011 and specific to each customer was sent via overnight courier mail. The letter described the affected product, issue and actions to take. Customers were instructed to check their inventory for the recalled products, cease use and distribution, and quarantine all affected product. Each customer received a custom copy of the letter with the specific product codes and lots that were shipped to them described in the Recall Acknowledgement and Inventory Return Form. Distributors were provided with instructions on how to handle product that was further distributed by them. The firm issued an updated Recall Notice dated May 31, 2011 to two consignees effected by the expanded recall. If you have any questions, please contact your local sales representative or Vygon's Customer Service Department at 1-800-473-5414, or by e-mail at customerservice@vygonus.com.
Details
- Recall number
- Z-2389-2011
- Initiated
- April 19, 2011
- Posted by FDA
- June 11, 2011
- Terminated
- June 18, 2012
- Product code
- LKB – Pad, Alcohol, Device Disinfectant
- Quantity
- 4,000 units
- Distribution
- United States (FL, MA, MD, and VA).
- Recalling firm
- Vygon Corporation, Montgomeryville, PA
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-2388-2011 | Churchill Medical Systems AMS-7080CP (Dressing Change Kit). The device intended use is a…Skin-Prep wipes purchased by Churchill (a Vygon company) for inclusion into Dressing Kits and PICC Insertion… | 04/19/2011Terminated |
| Z-2390-2011 | Churchill Medical Systems AMS-8431CP (PICC insertion Tray). The device intended use is a…Skin-Prep wipes purchased by Churchill (a Vygon company) for inclusion into Dressing Kits and PICC Insertion… | 04/19/2011Terminated |
| Z-2391-2011 | Churchill Medical Systems AMS-9189CP-1 (Dressing Change Kit) -Medium Glove. Churchill…Skin-Prep wipes purchased by Churchill (a Vygon company) for inclusion into Dressing Kits and PICC Insertion… | 04/19/2011Terminated |