Recall Z-2398-2026

Siemens Healthcare Diagnostics, Inc. · initiated May 7, 2026

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Siemens Dimension Tacrolimus Flex Reagent Cartridge (TAC), REF DF207, SMN 10700795; enzyme immunoassay

Codes / lots: Manufacturing Lot/Serial Numbers: FA6254, FA6260, FA6286; UDIs: (01)00842768035425(10)FA6254(17)20260911; (01)00842768035425(10)FA6260(17)20260917; (01)00842768035425(10)FA6286(17)20261013.

Reason for recall

A potential positive bias in quality control (QC) and patient sample results in three lots of the Dimension Tacrolimus (TAC) Flex reagent cartridge.

FDA-determined cause: Under Investigation by firm

Action taken

Siemens Healthineers issued an Urgent Medical Device Correction notice to its consignees on 5/12/2026 via FedEx. The notice described the issue, its impact to QC and patient results, and requested the following: "Customer Actions . Immediately discontinue use of and discard the affected kit lot numbers. . Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. . Review your inventory of these products to determine your laboratory s replacement needs and to provide information to Siemens Healthineers for required regulatory reporting. . Complete and return the Field Correction Effectiveness Check Form attached to this letter within (30) days. . Please retain this letter with your laboratory records and forward it to anyone who may have received or used this product." For questions, contact your Siemens Healthineers Customer Care Center at (312) 275-7795 or your local Siemens Healthineers technical support representative.

Details

Recall number
Z-2398-2026
Initiated
May 7, 2026
Posted by FDA
June 10, 2026
Product code
MLM – Enzyme Immunoassay, Tracrolimus
Quantity
2878 units
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Brazil, Croatia, Germany, Hungary, India, Italy, Japan, Korea, Lithuania, Pakistan, Poland, Saudi Arabia, Spain, South Africa, Sri Lanka, Turkey, United Arab Emirates.
Recalling firm
Siemens Healthcare Diagnostics, Inc., Newark, DE