Recall Z-2403-2016

OraSure Technologies, Inc. · initiated July 20, 2016

Class I Terminated

Class I – dangerous or defective products that could cause serious health problems or death

Product

OraQuick Ebola Rapid Antigen Test, Product Code 3001-2807; Intended for the qualitative detection of antigens for Ebola viruses.

Codes / lots: Product Code 3001-2807 and the lot number is 6648965. Expiration date October 31, 2016.

Reason for recall

Failed stability testing at 8 months. This lot produced intermittent false negative results for the positive kit control with devices stored at 30C conditions.

FDA-determined cause: Process control

Action taken

OraSure contacted the sole customer via email and mail (letter dated July 21, 2016) to request return of all the product from the affected lot.

Details

Recall number
Z-2403-2016
Initiated
July 20, 2016
Posted by FDA
August 17, 2016
Terminated
May 10, 2017
Quantity
3,075 units
Distribution
Distributed to one customer only in US state of Georgia.
Recalling firm
OraSure Technologies, Inc., Bethlehem, PA