Recall Z-2407-2020

Utak Laboratories Inc · initiated March 20, 2020

Class III Terminated

Class III – products unlikely to cause any adverse health reaction but that violate FDA labeling or manufacturing laws

Product

Utak NR Trace Elements Serum Control, 5 ml, REF 66816. The responsible firm on the label is UTAK LABORATORIES, INC., Valencia, CA. Control for in vitro diagnostics.

Codes / lots: Lot C2540

Reason for recall

The product has an incorrect reference value for cadmium and an incorrect expected range.

FDA-determined cause: Labeling Change Control

Action taken

The recalling firm issued a recall letter dated 3/19/2020 via email on 3/20/2020 requesting customers to destroy the incorrect IFU and replace it with the revised IFU.

Details

Recall number
Z-2407-2020
Initiated
March 20, 2020
Terminated
July 22, 2020
Product code
JJY – Multi-Analyte Controls, All Kinds (Assayed)
Quantity
1,395 vials
Distribution
Worldwide Distribution. US nationwide, Canada, Australia, France, Germany, Israel, Italy, New Zealand, Philippines, and United Kingdom.
Recalling firm
Utak Laboratories Inc, Valencia, CA