Recall Z-2414-2024

Beckman Coulter, Inc. · initiated June 4, 2024

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Dxl 9000 Access Immunoassay Analyzer, Catalog #C11137.

Codes / lots: Serial numbers: 300117, 300116, 300137, 300138, 300139, 300129, 300141, 300143, 300142, 300151, 300152, 300128, 300156, 300153, 300154, 300135, 300164, 300180, 300155, 300167, 300166, 300158, 300171, 300179, 300169, 300182, 300176, 300183, 300198, 300192, 300133, 300188, 300161, 300134, 300190, 300209, 300212, 300211, 300149, 300148, 300150, 300130, 300172, 300197, 300160, 300189, 300170, 300181, 300195, 300194, 300224, 300222, 300168, 300210, 300233, 300207, 300205, 300221, 300203, 300240, 300199, 300200, 300136, 300196, 300178, 300220, 300208, 300126, 300213, 300206, 300219, 300214, 300217, 300215, 300229, 300255, 300257, 300202, 300254, 300223, 300258, 300259, 300218, 300201, 300234, 300237, 300249, 300260, 300238, 300243, 300261, 300231, 300191, 300131, 300236, 300264, 300263, 300193, 300239, 300230, 300235, 300245, 300216, 300247, 300242, 300226, 300270, 300271, 300269, 300174, 300251, 300241, 300232, 300204, 300227, 300228, 300256, 300266, 300281, 300282, 300253, 300244, 300272, 300246, 300145, 300123, 300147, 300297, 300157, 300275, 300284, 300286, 300248, 300273, 300252, 300144, 300146, 300165, 300177, 300173, 300250, 300268, and 300140; UDI-DI 15099590732103.

Reason for recall

The analyzers were built with nonconforming USB cables made with brass contacts instead of specified gold contacts. Brass contacts may corrode over time, leading to loss of internal instrument communication and delayed patient results.

FDA-determined cause: Nonconforming Material/Component

Action taken

Letters dated 5/29/2024 were issued via email and postal mail on 6/4/2024. The letter described the issue, the impact, and the action which was that the consignee's Beckman Coulter service representative will contact the consignee to replace parts affected by the issue. The firm will replace all affected USB 2.0 cables. The information in the letter is to be shared with the consignees laboratory staff and if the device has been forwarded to another laboratory, the consignee is requested to send them a copy of the letter. The consignee is requested to respond to the letter within 10 days either via email or by manually completing the enclosed Response Form.

Details

Recall number
Z-2414-2024
Initiated
June 4, 2024
Posted by FDA
July 23, 2024
Product code
JJE – Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Quantity
143 devices
Distribution
Worldwide - US Nationwide distribution in the states of FL, IN, KS, NH, PA, and TX. The countries of Australia, Austria, Belgium, Brazil, Croatia, Czechia, Egypt, France, Germany, Hungary, Ireland, Israel, Italy, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, Spain, Switzerland, Taiwan, Province of China, and United Kingdom of Great Britain and Northern Ireland.
Recalling firm
Beckman Coulter, Inc., Chaska, MN

Official FDA recall record