Recall Z-2421-2010

Abbott Laboratories · initiated April 5, 2010

Terminated

Product

CELL-DYN Sapphire Probe Detector PCB Assembly (also called the Aspiration Bottom Sensor) used on the Cell-Dyn Sapphire Hematology Analyzer, Manufactured by Abbott Laboratories, Santa Clara, CA

Codes / lots: List 08H00-01, 08H00-03, PCB Assembly part number 8960167101

Reason for recall

Under certain conditions in closed mode of operations, the Aspiration Probe Detector PCB assembly may fail to operate properly and provide inaccurate results.

FDA-determined cause: Other

Action taken

Abbott Diagnostics Division of Santa Clara, CA initiated a field correction of the CELL-DYN Sapphire Probe Detector Assembly used on the CELL-DYN Sapphire Hemotology Analyzer, on April 5, 2010. The correction is being addressed through field service representatives and a mandatory Technical Service Bulletin (TSB). The TSB requires inspection, testing, and potential replacement of the affected assembly. The firm will not be issuing a seperate notification. For questions regarding this correction please contact (408) 982-4800, or fax (408) 982-4863.

Details

Recall number
Z-2421-2010
Initiated
April 5, 2010
Posted by FDA
September 14, 2010
Terminated
May 17, 2011
Product code
GKZ – Counter, Differential Cell
Quantity
926 units
Distribution
Product was distributed to 926 retailers throughout the US and to argentina, Australia, Brazil, Canada, Chile, China, Columbia, Germany, Hong King, Japan, Mexico, New Zealand, Panama, Paraguay, Peru, Singapore, South Korea, Taiwan and Venezuela.
Recalling firm
Abbott Laboratories, Santa Clara, CA

Official FDA recall record