Recall Z-2430-2008

Bio-Rad Laboratories · initiated July 29, 2008

Terminated

Product

Kallestad (TM) Mouse Kidney. Product labeled "BIO-RAD Kallestad (TM) Mouse Kidney Substrate... UNITED STATES, Bio-Rad Laboratories, Redmond, WA 98052... FRANCE, Bio-Rad Laboratories, 92430 Marnes La Coquette". Catalog numbers: 30445 - 8 wells, 48 tests and 30447 - 8 well, 192 tests.

Codes / lots: Lot # and exp: for catalog # 30445 - 951563, exp 11/30/08; 953560, exp 1/31/09; and 955471, exp 4/30/09; for catalog # 30447 - 951565, exp 11/30/08; 953561, exp 1/31/09; and 954607, exp 3/31/09.

Reason for recall

The Homogeneous Positive control was released with a titer of 1:256 rather than the titer value of 1:128 reported on the QC Report Label included with each test kit.

FDA-determined cause: Employee error

Action taken

On 7/25/08, the firm began to notify their customers about the recall via an Urgent Product Correction sent by DHL or FEDEX. Included with the customer letter was a revised QC Card. International subsidiaries were notified by e-mail with the customer letter and revised QC Card attached. The letter informed customers of the recall and directing them to the revised QC card which was attached to the letter. The instructions are to, "Please use this to replace the card in any kits remaining in your facility." If customers have any questions or require additional QC cards, they are to contact Technical Support.

Details

Recall number
Z-2430-2008
Initiated
July 29, 2008
Posted by FDA
September 19, 2008
Terminated
March 31, 2011
Product code
DBL – Multiple Autoantibodies, Indirect Immunofluorescent, Antigen, Control
Quantity
69 total kits
Distribution
Worldwide Distribution --- including USA and countries of AUSTRALIA, BRAZIL, CANADA, CHINA , FRANCE, GERMANY, HONG KONG, INDIA, ITALY, JAPAN, MEXICO, NEW ZEALAND, POLAND, SINGAPORE, SOUTH KOREA , TAIWAN, UNITED KINGDOM, and URUGUAY.
Recalling firm
Bio-Rad Laboratories, Redmond, WA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2428-2008 Kallestad (TM) HEp-2 Cell Line Substrate. Product labeled "BIO-RAD Kallestad (TM) HEp-2…The Homogeneous Positive control was released with a titer of 1:256 rather than the titer value of 1:128… 07/29/2008Terminated
Z-2429-2008 Kallestad (TM) Mouse Stomach/Kidney, catalog number 30443 - 8 wells, 48 tests. Product…The Homogeneous Positive control was released with a titer of 1:256 rather than the titer value of 1:128… 07/29/2008Terminated