Recall Z-2451-2023
Siemens Healthcare Diagnostics, Inc. · initiated July 25, 2023
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP), Material Number 11319151, in vitro diagnostic use in the quantitative determination of total protein in human urine and cerebrospinal fluid
Codes / lots: UDI-DI: 00630414279176; Lot Numbers: All lots
Reason for recall
There is a potential for ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein reagent carryover impacting Enzymatic Creatinine_2 (ECRE_2). Falsely depressed ECRE_2 results may be observed when the assay is processed after the UCFP test on ADVIA Chemistry systems.
FDA-determined cause: Under Investigation by firm
Action taken
An Urgent Medical Device Correction dated 7/25/23 was sent to customers. Actions to be Taken by the Customer Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. Perform the instructions provided in Additional Information. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. If you have received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. We apologize for the inconvenience this situation may cause. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.
Details
- Recall number
- Z-2451-2023
- Initiated
- July 25, 2023
- Posted by FDA
- August 24, 2023
- Product code
- JIT – Calibrator, Secondary
- Quantity
- 3067 units
- Distribution
- Worldwide - US Nationwide distribution in the states of CA, CO, CT, FL, GA, IA, MD, MI, NJ, NM, NY, OH, PA, SD, TX, UT, and VA. Global distribution.
- Recalling firm
- Siemens Healthcare Diagnostics, Inc., Tarrytown, NY