Recall Z-2455-2025
mo-Vis BVBA · initiated July 10, 2025
Class I – dangerous or defective products that could cause serious health problems or death
Product
Multi Joystick R-net. Electrical wheelchair component.
Codes / lots: Model No. P002-62; UDI-DI: 05407008320041; Serial No. 1000 to 1690.
Reason for recall
Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.
FDA-determined cause: Device Design
Action taken
An URGENT: MEDICAL DEVICE RECALL R-NET JOYSTICKS notification dated 07/10/2025 was sent to consignees instructing them to upgrade affected devices according to the provided Annex 2 or by emailing contact@mo-vis.com to coordinate return. The software upgrade of affected devices should be conducted during the next scheduled annual maintenance of the joystick. These actions should be verified. Questions can be directed via email to contact@mo-vis.com or by phone at +32 9 335 28 60.
Details
- Recall number
- Z-2455-2025
- Initiated
- July 10, 2025
- Posted by FDA
- September 2, 2025
- Product code
- KNN – Components, Wheelchair
- Quantity
- 3 units
- Distribution
- US Nationwide distribution in the state of TX.
- Recalling firm
- mo-Vis BVBA, Nevele
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-2454-2025 Class I | Micro Joystick R-net. Electrical wheelchair component.Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which… | 07/10/2025Open, Classified |
| Z-2456-2025 Class I | IDM-MICRO-R. Electrical wheelchair component.Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which… | 07/10/2025Open, Classified |
| Z-2457-2025 Class I | IDM-MULTI-R. Electrical wheelchair component.Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which… | 07/10/2025Open, Classified |
| Z-2458-2025 Class I | All-round Joystick R-net Light. Electrical wheelchair component.Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which… | 07/10/2025Open, Classified |
| Z-2459-2025 Class I | IDM-ARLITE-R. Electrical wheelchair component.Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which… | 07/10/2025Open, Classified |