Recall Z-2468-2025
American Contract Systems Inc · initiated July 14, 2025
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) SFCH CRANI PACK (PS 132432), MODEL/ITEM NUMBER UICR10W; 2) SFCH SHUNT PACK (PS 139798), MODEL/ITEM NUMBER UISH55V.
Codes / lots: 1) MODEL/ITEM NUMBER UICR10W: UDI/DI 00191072207241, Kit Lot Numbers: 42-7828711, 42-7980711, 42-8063211, 42-8124011; 2) MODEL/ITEM NUMBER UISH55V: UDI/DI 00191072235442, Kit Lot Numbers: 42-7799111, 42-7985211, 42-8065111, 42-8140711.
Reason for recall
During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.
FDA-determined cause: Under Investigation by firm
Action taken
Owens & Minor issued an "URGENT: MEDICAL DEVICE CORRECTION" notice to its consignees on 07/15/2025 via email. The notice explained the problem, potential risk to health, and requested the following: ACTIONS TO TAKE: " Discontinue use of all affected components. " Add warning labels to affected procedure trays to ensure the components are discarded. A labeling template is enclosed for printing labels which can be affixed on all affected kits in inventory. This label should be applied to each affected procedure tray and product case in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. o Customers (end-users): If you receive a product from a distributor, please contact the distributor or your sales representative with any questions you may have. o Distributors: If any of the affected procedure trays identified above were further distributed, please ensure all end-users are appropriately notified of this product correction and maintain records of effectiveness for your consignees. " Complete the attached Recall Response Form and return as soon as possible to GM-OMRA_RECALLS@owens-minor.com. Report any quality problems or adverse events associated with the products listed in this notification to complaints@owens-minor.com. If you have any questions, contact the Manager, Regulatory Affairs at 470-280-4277 or email to: GM-OMRA_RECALLS@owens-minor.com.
Details
- Recall number
- Z-2468-2025
- Initiated
- July 14, 2025
- Posted by FDA
- August 28, 2025
- Product code
- OJG – Neurological Tray
- Quantity
- 204 units
- Distribution
- US distribution to states of: IL and LA
- Recalling firm
- American Contract Systems Inc, Tiffin, OH
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-2464-2025 Class II | AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) HEART PACK -…During an internal investigation, ACS determined that the rationale for sterilization of certain components… | 07/14/2025Open, Classified |
| Z-2465-2025 Class II | AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) LAP CHOLE…During an internal investigation, ACS determined that the rationale for sterilization of certain components… | 07/14/2025Open, Classified |
| Z-2466-2025 Class II | AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) CYSTO…During an internal investigation, ACS determined that the rationale for sterilization of certain components… | 07/14/2025Open, Classified |
| Z-2467-2025 Class II | AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) A-C…During an internal investigation, ACS determined that the rationale for sterilization of certain components… | 07/14/2025Open, Classified |
| Z-2469-2025 Class II | AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) BORN ON…During an internal investigation, ACS determined that the rationale for sterilization of certain components… | 07/14/2025Open, Classified |
| Z-2470-2025 Class II | AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) LAPAROSCOPY…During an internal investigation, ACS determined that the rationale for sterilization of certain components… | 07/14/2025Open, Classified |
| Z-2471-2025 Class II | AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) ROBOT HYST…During an internal investigation, ACS determined that the rationale for sterilization of certain components… | 07/14/2025Open, Classified |
| Z-2472-2025 Class II | AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) ANTERIOR HIP…During an internal investigation, ACS determined that the rationale for sterilization of certain components… | 07/14/2025Open, Classified |