Recall Z-2470-2012

Alcon Research, LTD. · initiated May 29, 2012

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

This voluntary medical device correction is related to Alcon products that present a 23G nonvalved trocar cannula and infusion cannula together. These items are available as components within Accurus 23G TotalPLUS Pak, Accurus 23G Combined Procedure Packs (0.9mm and 1.1mm), Constellation 23G Total PLUS Pak, Constellation 23G Combined Pack (0.9 mm and 1.1 mm) and 23G Small Parts Kit. Single use 23G non-valved trocar cannulas are used to provide internal access to the ocular cavity during ophthalmic surgery. Single use infusion cannulas are used for the infusion of fluid or air during ophthalmic surgery to maintain intraocular pressure during posterior segment surgeries.

Codes / lots: 1) 23G TotalPLUS Pak: Accurus Standalone - 8065750828, 2) 23G Combined Pack 0.9mm: Accurus Standalone - 8065750814, 3) 23G Combined Pack 1.1mm: Accurus Standalone - 8065750816, 4)23G TotalPLUS Pak: Constellation Standalone - 8065751058, 5) 23G Combined Pack 0.9mm: Constellation Standalone -8065751077, 6) 23G Combined Pack 1.1mm: Constellation Standalone - 8065751065. 7) 23G Small Parts Kit - 8065750842. Complete list of lot numbers will be provided with recall recommendation.

Reason for recall

This medical device correction is related to Alcon products that present a 23G non-valved trocar cannula and infusion cannula together because of an issue regarding the loose fit between the 23G non-valved trocar cannula and infusion cannula that could potentially cause a detachment between the non-valved trocar cannula and infusion cannula during surgery.

FDA-determined cause: Device Design

Action taken

Alcon sent an Urgent a Voluntary Medical Device Correction letter on May 9, 2012. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed NOT to use the affected product, instead customers should use a 1-Count Valved Trocar Cannula and to complete and immediately return the Medical Device Correction Response Form to Alcon via fax at: (817) 916-9087 or email to: ReplacementCannulas@alconlabs.com. For questions customers should call 1-800-862-5266.

Details

Recall number
Z-2470-2012
Initiated
May 29, 2012
Posted by FDA
September 27, 2012
Terminated
February 12, 2013
Product code
NGY – Cannula, Trocar, Ophthalmic
Quantity
93,319 (30, 413 distributed domestically & 62,906 internationally)
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Saudi Arabia, United Kingdom, Greece, Singapore, Germany, Switzerland, Kuwait, United Arabian Emirates, and Austria.
Recalling firm
Alcon Research, LTD., Houston, TX

Official FDA recall record