Recall Z-2474-2020

Baxter Healthcare Corporation · initiated June 2, 2020

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Flow Coupler Monitor, GEM, Global Excellence Microsurgery, Rx Only

Codes / lots: Product Code: 5156-00000-011, All Serial numbers

Reason for recall

A firmware issue may cause the GEM1020M-2 Flow Coupler Monitor to intermittently shut down when a WiFi connection is not established.

FDA-determined cause: Device Design

Action taken

Synovis Micro Companies Alliance (MCA), Inc., a subsidiary of Baxter International Inc., sent a notification to healthcare providers dated, June 2, 2020 titled, "Urgent Medical Device Recall". The notification was sent to the user level for the GEM1020M-2 Flow Coupler Monitor due to a firmware issue which may cause the monitor to intermittently shut down when a WiFi connection is not established. If a shutdown occurs, the monitor can be manually powered back on and will resume normal operation. Synovis is requesting the return of the monitors to perform a firmware upgrade to resolve the issue. Customers should contact Synovis to arrange for product return for the firmware upgrade and replacement of the product. Customers should complete the Baxter customer reply form and return it to Baxter by scanning and e-mailing it to fca@baxter.com, even if customers do not have any inventory. Customers should return the response form promptly. If customers have distributed the product to other facilities or departments within the institution, a copy of the communication should be forwarded to them. Questions regarding the communication, contact Synovis at 800-510-3318 or 205-941-0111, between the hours of 8:00 am and 5:00 pm Central Time, Monday through Friday.

Details

Recall number
Z-2474-2020
Initiated
June 2, 2020
Product code
MVR – Device, Anastomotic, Microvascular
Quantity
60 units
Distribution
U.S. distribution only: VA, GA, MD, PA, DC, WI, LA, MN, FL, IN, MI, AZ, TX, MS, MA, NY, SC, WI, WA, MI, CA.
Recalling firm
Baxter Healthcare Corporation, Deerfield, IL