Recall Z-2476-2025

Diagnostica Stago, Inc. · initiated July 21, 2025

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

STA - Fibrinogen 5. Model/Catalog Number: 00674. The STA - Fibrinogen 5 kit is intended for use with the analyzers of the STA-R, STA Compact and STA Satellite families for the quantitative determination of fibrinogen levels in human citrated plasma by the clotting method of Clauss.

Codes / lots: Model/Catalog Number: 00674. UDI-DI: 13607450006749

Reason for recall

Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications.

FDA-determined cause: Mixed-up of materials/components

Action taken

Diagnostica Stago notified consignees on about 07/21/2025 via emailed letter titled URGENT: MEDICAL DEVICE RECALL (CORRECTION). Consignees were instructed to temporarily systematically run a quality control just before a batch of UFH or LMWH, if you run STA-LIQUID ANTI-Xa in UFH and LMWH applications on your STA SATELLITE analyzer. This request is being asked while Stago works to develop a permanent solution for the analyzer. Consignees were also requested to complete and return the Acknowledgement Form and to notify customers if affected units were transferred or further distributed.

Details

Recall number
Z-2476-2025
Initiated
July 21, 2025
Posted by FDA
August 28, 2025
Product code
KQJ – System, Fibrinogen Determination
Quantity
829,800 units (245,200 US)
Distribution
Worldwide - US Nationwide distribution in the states of AL, AZ, CA, CO, FL, GA, LA, MD, ME, MN, MO, MT, NC, OH, OK, OR, PA, PR, SD, VT, WI, WY and the countries of Algeria, Argentina, Austria, Brazil, Bulgaria, Canada, Chile, Colombia, Costa Rica, France, French Polynesie, Guadeloupe, India, Iran, Israel, Jordan, Korea, Kuwait, Lebanon, Malaysia, Mexico, Morocco, Netherlands, Oman, Paraguay…
Recalling firm
Diagnostica Stago, Inc., Parsippany, NJ

Official FDA recall record