Recall Z-2515-2025
Bio-Rad Laboratories, Inc. · initiated May 30, 2025
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Platelia Toxo IgM Catalog Number 26211 UDI Code: 03610520005552 The Platelia Toxo IgM assay is a semi-quantitative immunocapture enzyme immunoassay kit for the detection of IgM antibodies to Toxoplasma gondii in human serum, plasma or cord blood specimens.
Codes / lots: Catalog Number: 26211 UDI-DI code: 03610520005552 Batch/Lot Number: 4L0054
Reason for recall
Due to a risk of false positive results that could lead to unnecessary medical treatment.
FDA-determined cause: Process control
Action taken
On 05/30/2025, the firm sent via email an "URGENT MEDICAL DEVICE NOTIFICATION" to customers informing them that, as a result of an internal investigation, it was revealed that samples taken from a frozen negative control panel tested unexpectedly interpreted as equivocal or positive results. Customer are instructed to: . As the issue can alter product performance of the affected lots by reducing specificity and increasing the number of equivocal results, interpret positive results with caution and contact Bio-Rad Technical Support if you observe an unusual rate of false positive results. . Refer to the test limitations, as outlined in Section 9 Limitations of the Procedure, of the IFU 0002043 version 2023/11: Diagnosis of recent infection by T. gondii can only be established on the basis of a combination of clinical and serological data. The result of a single serum sample does not constitute sufficient proof for diagnosis of recent infection . . As noted in Section 8 Interpretation of Results, positive IgM results should be followed by confirmatory steps in accordance with CDC guidelines before initiating or continuing treatment. For questions or assistance, contact Bio-Rad Technical Support at 1-800-224-6723 or by email at TechSupportUSSD-Redmond@Bio-Rad.com
Details
- Recall number
- Z-2515-2025
- Initiated
- May 30, 2025
- Posted by FDA
- September 4, 2025
- Product code
- LGD – Enzyme Linked Immunoabsorbent Assay, Toxoplasma Gondii
- Quantity
- 35 kits
- Distribution
- U.S. Nationwide distribution in the states of CA, CO, FL, NC and NV.
- Recalling firm
- Bio-Rad Laboratories, Inc., Woodinville, WA