Recall Z-2518-2023

Aesculap Implant Systems LLC · initiated July 20, 2023

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

For intraventricular indications mainly used for operations within the brain ventricles. Product Code: FF399R

Codes / lots: UDI-DI: 04038653065547 Lot Numbers: 52750146 52753381, 52756472, 52768317, 52805655, 52812596, 52816570, 52823878, 52830872. Serial Numbers: 4281,4282, 4284,4288, 4289,4291, 4308,4309, 4310, 4312, 4313,4336, 4340,4341, 4342,4402, 4411, 4412, 4413,4414, 4422,4439, 4447,4449, 4450

Reason for recall

Trocar manufactured with the shaft too long and does not meet manufacturing specifications. As a result, when used with the 30 neuroendoscope, a portion of the visual field is obscured.

FDA-determined cause: Process change control

Action taken

Aesculap issued Urgent Medical Device Recall Notification Letter via USPS Certified Mail 20.JUL.2023. Letter states reason for recall, health risk and action to take: 1. Review the Recall Notification in its entirety and ensure that all users in your organization of the above-mentioned product, and all concerned persons are informed about this voluntary product recall. 2. Your Aesculap Sales Representative will be in contact with you to review your current inventory, aid in evaluating your devices and assist you in completing the attached Product Acknowledgement Form. 3. Do not destroy any affected product. 4. Utilizing the attached form, you and your Sales Representative will record the total number of individual impacted pieces. If you have no inventory remaining, a zero (0) will be entered on the form. 5. Return the completed "Product Acknowledgement Form" to Aesculap, Inc. Product Quality Excellence department by emailing the form to PA_QualityAssurance.BBMUS_Service@bbraunusa.com or faxing the form to (610) 849-1197 after your inventory has been evaluated by your Sales Representative, even if the total inventory in your possession is zero (0). 6. Once Aesculap, Inc. receives your "Product Acknowledgement Form", a customer support representative will contact you with instructions on how to return any impacted pieces in your possession.

Details

Recall number
Z-2518-2023
Initiated
July 20, 2023
Posted by FDA
September 6, 2023
Product code
GWG – Endoscope, Neurological
Quantity
25 units
Distribution
US Nationwide distribution in the states of FL, ID, IL, MD, NC, NY, OR, RI, TX.
Recalling firm
Aesculap Implant Systems LLC, Center Valley, PA