Recall Z-2524-2024

American Contract Systems, Inc. · initiated May 6, 2024

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

American Contract Systems Spine Pack convenience kit

Codes / lots: UDI-DI 00191072202499; Lot 980241; Exp. Date 1/10/2025.

Reason for recall

Product released with insufficient aeration time impacting the kits sterilized on the same cycle or in the same aeration chamber

FDA-determined cause: Under Investigation by firm

Action taken

The recalling firm issued recall letters on and dated 5/6/2024 via email or overnight mail. The letter explained the event details, product involved, risk to health, and actions to take. The actions included (1) review your inventory and segregate and quarantine all affected trays. If any of the trays have been further distributed, ensure all end users are notified and maintain records of effectiveness; (2) discard all affected kits and document on the enclosed Response Form; and (3) complete the enclosed Response Form and return it as soon as possible.

Details

Recall number
Z-2524-2024
Initiated
May 6, 2024
Posted by FDA
August 7, 2024
Product code
OJG – Neurological Tray
Quantity
3
Distribution
US Nationwide distribution in the states of CA, KS, ME, MN, MO, NE, and SD.
Recalling firm
American Contract Systems, Inc., Kansas City, MO

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2517-2024 Class IIAmerican Contract Systems Hand Pack convenience kitProduct released with insufficient aeration time impacting the kits sterilized on the same cycle or in the… 05/06/2024Open, Classified
Z-2518-2024 Class IIAmerican Contract Systems Gyn Laparoscopy convenience kitProduct released with insufficient aeration time impacting the kits sterilized on the same cycle or in the… 05/06/2024Open, Classified
Z-2519-2024 Class IIAmerican Contract Systems Laparoscopy Litho Std Pack convenience kit and American…Product released with insufficient aeration time impacting the kits sterilized on the same cycle or in the… 05/06/2024Open, Classified
Z-2520-2024 Class IIAmerican Contract Systems Special Procedure Pack convenience kitProduct released with insufficient aeration time impacting the kits sterilized on the same cycle or in the… 05/06/2024Open, Classified
Z-2521-2024 Class IIAmerican Contract Systems Extremity convenience kit, American Contract Systems Hand Pack…Product released with insufficient aeration time impacting the kits sterilized on the same cycle or in the… 05/06/2024Open, Classified
Z-2522-2024 Class IIAmerican Contract Systems Cataract Pack convenience kitProduct released with insufficient aeration time impacting the kits sterilized on the same cycle or in the… 05/06/2024Open, Classified
Z-2523-2024 Class IIAmerican Contract Systems Biopsy Drape Pack convenience kitProduct released with insufficient aeration time impacting the kits sterilized on the same cycle or in the… 05/06/2024Open, Classified