Recall Z-2531-2020

Philips North America, LLC · initiated June 8, 2020

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Philips M3015A Microstream CO2 Extension Model No. 862393 -used with an lntelliVue X3 Patient Monitor (Model No. 867030) or a lntelliVue MMX Multi-Measurement Module (Model No. 867036)

Codes / lots: Units with this hardware version can be identified by the following serial number prefix: SN prefix: DE435xxxxx (first five digits)

Reason for recall

Affected M3015A extensions are incompatible with the lntelliVue X3 Patient Monitor (Model No. 867030) and lntelliVue MMX multi-measurement server (Model No. 867036).If they are connected to a host monitor, communication with the X3 or MMX could be disrupted, resulting in a loss of monitoring on the host monitor.

FDA-determined cause: Device Design

Action taken

Philips issued Customer Information Medical Device Correction (CIL86201894B) via priority mail on 6/8/20. Philips Market Organizations are responsible for distributing the letters outside of the U.S. Letter states reason for recall, health risk and action to take: To check if this issue affects your M3015A, compare the serial number of your device with the affected serial numbers identified; Use affected M3015A extensions only together with an lntelliVue X2 or MMS.; Do not combine an affected M3015A with an lntelliVue X3 or MMX; Philips Healthcare is distributing with this letter an addendum to the Instructions for Use for patient monitors that include directions for the use of the M3015A (MX100/X3, MX400-800 and MX750/MX850). If you need any further information or support concerning this issue, please contact your local Philips representative or Philips Customer Support at 1-800-722-9377.

Details

Recall number
Z-2531-2020
Initiated
June 8, 2020
Terminated
January 9, 2023
Product code
DSI – Detector And Alarm, Arrhythmia
Quantity
46,597 Units WW
Distribution
Nationwide Foreign: Angola Angola Argentina Australia Austria Bahrain Belarus Belgium Bolivia Brazil Bulgaria Canada Chile China Colombia Costa Rica Croatia Cyprus Czech Republic Denmark Egypt Estonia Finland France French Guiana Germany Ghana Gibraltar Greece Greenland Guadeloupe Guatemala Hong Kong Hungary Iceland India Indonesia Iran Ireland Israel Italy Japan Jordan Kazakhstan Kenya Korea…
Recalling firm
Philips North America, LLC, Andover, MA

Official FDA recall record