Recall Z-2539-2020

OrthoPediatrics Corp · initiated June 5, 2020

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

OrthoPediatrics ACL Reconstruction System ShieldLoc 7mm implant component model number 10-1008-4070 - Product Usage: is intended for fixation of tendons and ligaments during orthopedic reconstruction procedures such as Anterior Cruciate Ligament (ACL) Reconstruction.

Codes / lots: UDI: (01)00841132122211(17)240722(10)214320319A02 Lot#: 214320319A02

Reason for recall

ShieldLoc implant component of the ACL Reconstruction Systemlots may be labeled with the incorrect size. Product labeled as 10-1008-4070 was found to have 10-1008-4060 product in the package. Product labeled as 10-1008-4060 was found to have 10-1008-4070 product in the package.

FDA-determined cause: Under Investigation by firm

Action taken

OrthoPediatrics Corporation issued Urgent Medical Device Correction notices dated June 5, 2020 to customers via email. Customers are advised to return affected product to OrthoPediatrics Corporation. Replacement product will be provided. Please contact customer service to arrange return of the affected product. No additional corrective actions need to be taken for product already implanted. Please complete and return the attached response form confirming your understanding of these instructions. Any questions regarding this recall can be directed to Kriss Anderson, VP of Quality Systems and Regulatory Affairs, at OrthoPediatrics at kanderson@orthopediatrics.com or 574-267-0897.

Details

Recall number
Z-2539-2020
Initiated
June 5, 2020
Terminated
May 25, 2021
Product code
MBI – Fastener, Fixation, Nondegradable, Soft Tissue
Quantity
11 units
Distribution
US Nationwide distribution.
Recalling firm
OrthoPediatrics Corp, Warsaw, IN

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2538-2020 Class IIOrthoPediatrics ACL Reconstruction System ShieldLoc 6mm implant component model number…ShieldLoc implant component of the ACL Reconstruction Systemlots may be labeled with the incorrect size.… 06/05/2020Terminated