Recall Z-2540-2020

FHC, Inc. · initiated June 17, 2020

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

FHC Guideline 4000 5.0 GL5 Main Processing Unit - Product Usage: intended to record and stimulate electrophysiological activity, as well as aid in the accurate placement of electrodes and other instruments. Catalogue Number: C0215 UDI: 00873263006740

Codes / lots: Serial Numbers: 45695, 48871, 48873, 48874, 46195, 46198

Reason for recall

An electrical sub-circuit was not properly connected. Its purpose is to eliminate the possibility of interference with proper performance of the Guideline 5 Main Processing Unit, if an ESD event (greater than 6kV) were to occur.

FDA-determined cause: Component change control

Action taken

FHC issued Urgent Medical Device Recall letter dated 6/17/20 via USPS certified return receipt,stating reason for recall, health risk and action to take: FHC, Inc. field representatives will be contacting you to return the effected product and issue a replacement. In the interim, we advise you to stop use of this product. Should you have any further questions regarding this matter you may contact me, by email at kmoeykens@fh-co.com or by telephone at 207-666-5425.

Details

Recall number
Z-2540-2020
Initiated
June 17, 2020
Terminated
November 2, 2020
Product code
GZL – Electrode, Depth
Quantity
6 units
Distribution
US Nationwide distribution including in the states of CA, NC, TN, TX, WA.
Recalling firm
FHC, Inc., Bowdoin, ME

Official FDA recall record