Recall Z-2545-2024

Beckman Coulter, Inc. · initiated June 14, 2024

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Discrete photometric chemistry analyzer for clinical use.

Codes / lots: Catalog Number: 81600N, C11907; UDI: 150995901; Serial Numbers: 574415, 574440, 574484,574416, 574441, 574485, 574417, 574442, 574486, 574418, 574443, 574488,574419, 574444, 574489,574420, 574445, 574490, 574421, 574446, 574491, 574422, 574447 574494,574423, 574448, 574497,574424, 574454, 574502, 574425, 574455, 574503, 574426, 574456 574504,574427, 574457, 574505,574428, 574459, 574506, 574429, 574469, 574507, 574430, 574470 574509,574432, 574471, 574513, 574433, 574472, 574514, 574434, 574477, 574515, 574435, 574479

Reason for recall

Beckman Coulter, Inc. has identified a rise in reagent carousel motion errors in Access 2 Immunoassay System Analyzers that appear to be caused by misaligned index wheels housed within the reagent carousel module. Customers affected by this issue may observe an increase in reagent carousel motion errors or QSMC errors that occur while unloading or loading reagent packs during system initialization and sample processing, which causes a non-operational system status and cancels tests; and patient results may be delayed if these errors cancel test processing.

FDA-determined cause: Under Investigation by firm

Action taken

The recall notification was sent to consignees during the week of 06/14/2024 via email and postal mail. Consignees were instructed to follow the troubleshooting instructions in section 6.4 of the Access 2 Reference Manual and contact the Beckman Coulter representative if the laboratory has experienced repeated reagent carousel motion or QSMC errors. Consignees are asked to respond within 10 days indicating they have received the important communication either via by email or by returning the response form.

Details

Recall number
Z-2545-2024
Initiated
June 14, 2024
Posted by FDA
August 9, 2024
Product code
JJE – Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Quantity
59 units
Distribution
Worldwide - US Nationwide distribution in the states of NE, MT, LA, TX, NY, CA, MI, AL, IL the countries of Austria, Belgium, Georgia, India, Indonesia, Italy, Korea, Malaysia, Myanmar, Nigeria, Switzerland, Vietnam.
Recalling firm
Beckman Coulter, Inc., Chaska, MN

Official FDA recall record