Recall Z-2546-2015

Covidien LLC · initiated August 19, 2015

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

MediTrace Cadence Adult Zoll, Preconnect Defibrillation Electrode, Nonsterile Item Code:22770PC

Codes / lots: Lot Numbers: 516313X, 519124X, 513426X

Reason for recall

Multi-Function Defibrillation Electrodes, Pre connect- potential for damage to the wire insulation may result in a potentially increased risk for reduced or no patient therapy and patient and/or clinician burns

FDA-determined cause: Process control

Action taken

Medtonic/Covidien issued recall letter on August 19, 2015. Customers were notified via Federal Express and the letter instructs customers to discontinue use of the product and return all inventory to Medtronic (Covidien), 2 Ludlow Park Drive, Chicopee, MA 01022, Attention: Field Action Returns. Contact your Medtronic representative or Customer Service at (800)-882-5878.

Details

Recall number
Z-2546-2015
Initiated
August 19, 2015
Posted by FDA
September 2, 2015
Terminated
May 24, 2017
Product code
DRO – Pacemaker, Cardiac, External Transcutaneous (Non-Invasive)
Quantity
114,000 sets
Distribution
Nationwide Foreign: Canada Australia Denmark Germany New Zealand
Recalling firm
Covidien LLC, North Haven, CT

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2545-2015 Class IIMediTrace Cadence Adult Multi-Function Defibrillation Electrodes Item Code:22550PCMulti-Function Defibrillation Electrodes, Pre connect- potential for damage to the wire insulation may… 08/19/2015Terminated
Z-2547-2015 Class IIPhysio-Control QuikCombo Pacing/Defibrillation/ECG Electrodes RediPak Preconnect System…Multi-Function Defibrillation Electrodes, Pre connect- potential for damage to the wire insulation may… 08/19/2015Terminated