Recall Z-2550-2025
Boston Scientific Corporation · initiated August 18, 2025
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
ICEfx Cryoablation System, portable helium-free desktop cryoablation unit, Material Numbers (UPN) FPRCH8000, FPRCH8000-02, H749396118000
Codes / lots: FPRCH8000 UDI-DI: 00859019006370, Serial Numbers: ICP0305, ICP0369, ICP0338; FPRCH8000-02 UDI-DI: 00859019006417, Serial Numbers: 86643865, 108415728, 88589050, 108470986, 89392275, 108612524, 107799735, ICP0299, 107887794, 3779, 108254011, IF0096 108312221, 104733618; H749396118000 UDI-DI: 00191506014995, Serial Numbers: 106066356, 107851990
Reason for recall
Certain desiccant tube subassemblies were built with end caps that were not correctly tightened.
FDA-determined cause: Process control
Action taken
An Urgent Medical Device Correction dated 8/18/25 was sent to customers. Customers should cease use of the systems listed in Table 1 and segregate them until a Boston Scientific representative has corrected this issue. A Boston Scientific Field Service Engineer will contact your facility to schedule the correction of your system. For patients in whom an impacted device has already been used, there are no specific recommendations beyond the local standard post-procedure care. If you require additional assistance or more information regarding this communication, please contact your local Boston Scientific representative. The Acknowledgement Form enclosed with this product correction must be completed and returned even if you do not have any affected units. An Urgent Medical Device Removal notification letter dated 8/18/25 was sent to a customer regarding the PM Kit ICEFX. Instructions: 1. Further distribution or use of any devices (Table 1) affected by this removal should cease immediately. Do NOT use affected devices and remove those devices from your facility s inventory. Segregate the units in a secure place until they can be returned to Boston Scientific. 2. Immediately post this information in a visible location near the affected devices to ensure it is readily accessible to all handlers and users of the device. 3. Forward this notice to any healthcare professional from your organization and to any facilities where affected devices have been transferred, including hospitals or sites within your network. 4. Complete and return the enclosed Reply Verification Tracking Form per the enclosed instructions on page three. 5. Return affected devices. If you require additional assistance or more information regarding this communication, please contact your local Boston Scientific representative.
Details
- Recall number
- Z-2550-2025
- Initiated
- August 18, 2025
- Posted by FDA
- September 11, 2025
- Product code
- GEH – Unit, Cryosurgical, Accessories
- Quantity
- 18 units
- Distribution
- Worldwide - US Nationwide and the countries of Canada, France, Germany and Italy. (Preventive Maintenance Kit distributed to Canada only.)
- Recalling firm
- Boston Scientific Corporation, Maple Grove, MN
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-2551-2025 Class II | Preventive Maintenance (PM) Kit ICEFX, Material Number (UPN) SPRPM8000Certain desiccant tube subassemblies were built with end caps that were not correctly tightened. | 08/18/2025Open, Classified |