Recall Z-2552-2019
Sentinel CH SpA · initiated June 28, 2019
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Alinity c Lithium Reagent Kit- intended for the quantitation of lithium in serum or plasma using respectively the ARCHITECT c Systems and the Alinity c analyzer List Number: 08P5320
Codes / lots: Lot Number/UDI: 80553Y600/ (01)00380740136277 (17)190731(10)80553Y600; 80444Y600/ (01)00380740136277 (17)190731(10)80444Y600; 80625Y600/ (01)00380740136277 (17)191130(10)80625Y600; 90224Y600/ (01)0038074013627 (17)200229(10)90224Y600
Reason for recall
Added an additional SmartWash parameter for the Lithium and Alinity c Lithium Reagent Kit, catalogue number 8L25-30 and 08P5320 to prevent the potential for carryover from the ARCHITECT Lactate Dehydrogenase (LDH) reagent (LN 2P56) may result infalsely elevated lithium patient results
FDA-determined cause: Under Investigation by firm
Action taken
Sentinel issued Urgent Field Safety Notice letter on 6/28/19 via email to the distributor. The Distributor Abbott notified customers by letter Urgent Field Safety Notice Product Correction dated July 11, 2019 advising of the problem, health risk and action to take: If the LDH reagent is installed on your instrument, please follow the instructions 1. Update the Lithium (LN 8L25-30) assay parameters to add the new SmartWash. For additional information, see the ARCHITECT System Operations Manual, Section 2, Configure the SmartWash settings (c System. Install the Alinit y c Lithium assay file version 6 to implement the new LDH SmartWash. Customer Service at 18774ABBOTT (available 24 hours a day, 7 days a week). Customers outside the U.S., please contact your local area Customer Service
Details
- Recall number
- Z-2552-2019
- Initiated
- June 28, 2019
- Terminated
- July 24, 2020
- Product code
- NDW – Assay, Porphyrin, Spectrophotometry, Lithium
- Quantity
- 1217 kits
- Distribution
- Nationwide Foreign: ARGENTINA AUSTRALIA AUSTRIA AZERBAIJAN BANGLADESH BELARUS BELGIUM BOSNIA AND HERZEGOVI BRAZIL BRUNEI CANADA CHILE COLOMBIA COSTA RICA CROATIA CZECH REPUBLIC DENMARK FAROE ISLANDS FINLAND FRANCE GERMANY GREECE GREENLAND HONG KONG IRELAND ISRAEL ITALY JORDAN KUWAIT LEBANON LUXEMBOURG MALAYSIA MALDIVES MEXICO MOROCCO NETHERLANDS NEW ZEALAND NIGERIA NORWAY OMAN PAKISTAN PANAMA…
- Recalling firm
- Sentinel CH SpA, Milan