Recall Z-2554-2026

IPG Medical Corporation · initiated May 5, 2026

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Thuvera Laser Console. Software Version 2.2. Portable Thulium Fiber Laser.

Codes / lots: Model No. MD2300F-001507; UDI-DI 00810071230226; Serial No. MDUF250342, MDUF250400, MDUF250429, MDUF250060, MDUF250061, MDUF250067, MDUF250068, MDUF250069, MDUF250071, MDUF250076, MDUF250083, MDUF250086, MDUF250091, MDUF250092, MDUF250104, MDUF250131, MDUF250133, MDUF250134, MDUF250159, MDUF250160, MDUF250199, MDUF250200, MDUF250202, MDUF250203, MDUF250204, MDUF250205, MDUF250240, MDUF250241, MDUF250242, MDUF250278, MDUF250297, MDUF250299, MDUF250312, MDUF250314, MDUF250325, MDUF250340, MDUF250341, MDUF250351, MDUF250352, MDUF250355, MDUF250367, MDUF250368, MDUF250370, MDUF250380, MDUF250384, MDUF250385, MDUF250395, MDUF250399, MDUF250409, MDUF250410, MDUF250427, MDUF250430, MDUF250431, MDUF250433, MDUF260009, MDUF260010, MDUF260011, MDUF260017, MDUF260019, MDUF260023, MDUF260024, MDUF260025, MDUF260026, MDUF260027, MDUF260032, MDUF260043, MDUF260067, MDUF260068, MDUF260070, MDUF260072, MDUF260077, MDUF260079, MDUF260094, MDUF260095, MDUF260096, MDUF260097, MDUF260098, MDUF260099, MDUF260100, MDUF260101, MDUF260103, MDUF260104, MDUF260105, MDUF260106, MDUF260107, MDUF260108, MDUF260109, MDUF260112, MDUF260115, MDUF250014, MDUF250015, MDUF250016, MDUF250050, MDUF250051…

Reason for recall

Software anomaly that causes a false display of error code 5018.

FDA-determined cause: Device Design

Action taken

An URGENT MEDICAL DEVICE RECALL notice dated 05/05/2026 was emailed to IPG Medical's consignee notifying them of this recall correction to address Error 5018 and Error 5007/5023 displayed on monitors with software version 2.2. The notification provides instructions to consignees to work around the error messages until a software upgrade is deployed at a later date. The notification instructs consignees that their staff may correct units onsite via software upgrade or units may be returned to IPG Medical for a factory correction when the update is available. A corrected IFU will be provided to all customers. Questions can be directed via phone to 774-633-5903 or koconnell@ipgmedical.com.

Details

Recall number
Z-2554-2026
Initiated
May 5, 2026
Posted by FDA
June 24, 2026
Product code
GEX – Powered Laser Surgical Instrument
Quantity
198 units
Distribution
US Nationwide distribution in the state of MA.
Recalling firm
IPG Medical Corporation, Marlborough, MA

Official FDA recall record