Recall Z-2558-2011

OPTI Medical Systems, Inc · initiated March 15, 2011

Terminated

Product

OPTI LION E-Plus Cassette, Model BP7507, Opti Medical (25 Cassettes with Samples Probes per box). Cassettes are consumables used in the OPTI LION Electrolyte Analyzer intended to be used for the measurement of sodium potassium, chloride, ionized calcium and pH in samples of whole blood, serum and plasma. . Cassettes contain one-time use sensors that are used with the OPTI-LION analyzer to perform in-vitro measurements of electrolytes. OPTI LION E-Plus Cassettes are supplied in boxes of 25 and may only be used with the OPTI LION Electrolyte Analyzer.

Codes / lots: Lot numbers 038811 and 102812

Reason for recall

The values reported by the chloride sensors may report erroneous values below specification on aqueous controls and potentially with patient samples of whole blood and serum/plasma.

FDA-determined cause: Nonconforming Material/Component

Action taken

OPTIMedical Systems sent an "URGENT: MEDICAL DEVICE RECALL" letter dated March 15, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. Customers were notified by letter via e-mail and fax. Customers were instructed to discontinue use of the product and to return any unused product to OPTI Medical Systems, Inc. Customers outside the U.S. were instructed to destroy unused product and certify the destruction on the enclosed Recall Response Card. The Response Card was to be completed and returned to OPTI Medical Customer Service via fax. OPTI Medical Systems will replace the cassettes they have with product from another lot. Customers were instructed to contact the Technical Support Department at 1-800-490-6784 for questions relating to this notice.

Details

Recall number
Z-2558-2011
Initiated
March 15, 2011
Posted by FDA
June 14, 2011
Terminated
August 8, 2011
Product code
CGZ – Electrode, Ion-Specific, Chloride
Quantity
86 boxes (25 units (per box)
Distribution
Worldwide Distribution-- USA (nationwide) including states of CA and VA, and countries of India, Lebanon, Philippines, Sweden, Venezuela and Vietnam.
Recalling firm
OPTI Medical Systems, Inc, Roswell, GA

Official FDA recall record