Recall Z-2567-2025

MEDLINE INDUSTRIES, LP - Northfield · initiated July 25, 2025

Class I Open, Classified

Class I – dangerous or defective products that could cause serious health problems or death

Product

Medline medical convenience kits labeled as: 1. NEWBORN RESUSCITATION KIT, Kit Number ACC010381C; 2. DRAWER 5 - POSITION 3, Kit Number ACC010532; 3. DRAWER #4C ADULT, Kit Number ACC010671; 4. DRAWER #2, Kit Number ACC010728.

Codes / lots: 1. ACC010381C: UDI/DI 10193489922127 (EA), 40193489922128 (CS), Kit Lot Numbers: 2. ACC010532: UDI/DI 10193489902358 (EA), 40193489902359 (CS), Kit Lot Numbers: 3. ACC010671: UDI/DI 10195327351090 (EA), 40195327351091 (CS), Kit Lot Numbers: 4. ACC010728: UDI/DI 10195327506544 (EA), 40195327506545 (CS), Kit Lot Numbers:

Reason for recall

Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing Flexicare BritePro Solo Fiber Optic Laryngoscope Handles due to the failure of the light to illuminate as intended.

FDA-determined cause: Component design/selection

Action taken

Medline Industries issued a MEDICAL DEVICE RECALL notice to it consignees on 07/25/2025 via email and USPS first class mail. The notice explained the issue with the component within the kit, potential risk to the patient, and requested the following: REQUIRED ACTION: 1.Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2.Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.comRecall Reference #: R-25-158-FGX1 Recall Code: 3.Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4.If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.

Details

Recall number
Z-2567-2025
Initiated
July 25, 2025
Posted by FDA
September 19, 2025
Product code
LRO – General Surgery Tray
Quantity
596 units
Distribution
US distribution to: AL, CA, DE, OR, TN, TX
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield, Northfield, IL

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2566-2025 Class IMedline medical convenience kits labeled as: 1. ADULT INTUBATION TRAY, Kit Number…Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing… 07/25/2025Open, Classified