Recall Z-2570-2011
Roche Diagnostics Operations, Inc. · initiated May 18, 2011
Product
ACCU-CHECK PERFORMA STRIP NO. 05987288031, 100 BANDELETTES REACTIVES, ROCHE DIAGNOSTICS GMBH, WWW.ACCUCHEK.COM glucose monitoring test strips.
Codes / lots: Lot Number 470049, expiration date 12/31/2011.
Reason for recall
Erroneous results: Internal investigations revealed that the test strips exhibited a performance issue that can cause test strip errors or a bias in test result that exceed the tolerance range according to EN ISO 15197.
FDA-determined cause: Component design/selection
Action taken
Direct accounts were notified by a Roche Diagnostics press release on 5/18/2011 that included the product's name, lot number, and information to withdraw the remaining inventory of lot 470049 from storage
Details
- Recall number
- Z-2570-2011
- Initiated
- May 18, 2011
- Posted by FDA
- June 21, 2011
- Terminated
- October 25, 2011
- Product code
- NBW – System, Test, Blood Glucose, Over The Counter
- Quantity
- 27,019 units
- Distribution
- No US distribution, in France Only.
- Recalling firm
- Roche Diagnostics Operations, Inc., Indianapolis, IN