Recall Z-2572-2026
Surgify Medical OY · initiated April 30, 2026
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Surgify Halo, 5.4 mm, Long, Model/Catalog Number: 54.140.SHD.H1; drills, burrs, trephines & accessories (simple, powered)
Codes / lots: UDI: 06429811532137 ; All lots until the IFU update has been implemented
Reason for recall
Potential for burr breakage, occurring during bi-portal endoscopic spinal surgery (BESS).
FDA-determined cause: Under Investigation by firm
Action taken
Surgify issued a Medical Device Correction Notice - Surgify Halo - Restriction of Use notice to its consignees on 4/30/2026 via email. The notice explained the issue, potential risk, and requested the following: "What steps must the user take? Please review your current practice and ensure the device is only used in accordance with the updated safety information. Instruction to Customers: The customers have been instructed to do the following: - Do not use this device in endoscopic surgical procedures. If endoscopic access is required, an alternative instrument should be selected. Specifically: . Do not use this device in endoscopic surgical procedures. If endoscopic access is required, an alternative instrument should be selected. . Ensure that all relevant clinical staff and surgical team members are made aware of this notice. . Please retain this notice for your records and attach it to the relevant product documentation. Please acknowledge receipt of this Medical Device Correction Notice by replying to raqa@surgifymedical.com within 30 business days, confirming that the notice has been read, understood, and communicated to all relevant users within your organization. Please include your facility name, your name, and your role. If you have previously used the device in endoscopic procedures without incident, no patient follow-up is required; however, this application should now be discontinued in line with the updated safety information." For questions regarding this matter, please contact us by phone at +358 105 176 310 or via email at support@surgifymedical.com.
Details
- Recall number
- Z-2572-2026
- Initiated
- April 30, 2026
- Posted by FDA
- June 24, 2026
- Product code
- HBE – Drills, Burrs, Trephines & Accessories (Simple, Powered)
- Distribution
- US Nationwide distribution in the states of MN, MA, IN, LA, CA, NY, FL.
- Recalling firm
- Surgify Medical OY, Espoo
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-2555-2026 Class II | Surgify Halo, 3.0 mm, Extendable, Model/Catalog Number: 30.000.SEE.U2; drills, burrs…Potential for burr breakage, occurring during bi-portal endoscopic spinal surgery (BESS). | 04/30/2026Open, Classified |
| Z-2556-2026 Class II | Surgify Halo, 3.0 mm, Short, Model/Catalog Number: 30.070.NVG.U2; drills, burrs…Potential for burr breakage, occurring during bi-portal endoscopic spinal surgery (BESS). | 04/30/2026Open, Classified |
| Z-2557-2026 Class II | Surgify Halo, 3.0 mm, Long, Model/Catalog Number: 30.125.NVG.U2; drills, burrs…Potential for burr breakage, occurring during bi-portal endoscopic spinal surgery (BESS). | 04/30/2026Open, Classified |
| Z-2558-2026 Class II | Surgify Halo, 4.0 mm, Extendable, Model/Catalog Number: 40.000.SEE.H1; drills, burrs…Potential for burr breakage, occurring during bi-portal endoscopic spinal surgery (BESS). | 04/30/2026Open, Classified |
| Z-2559-2026 Class II | Surgify Halo, 4.0 mm, Extendable, Model/Catalog Number: 40.000.SEE.U1; drills, burrs…Potential for burr breakage, occurring during bi-portal endoscopic spinal surgery (BESS). | 04/30/2026Open, Classified |
| Z-2560-2026 Class II | Surgify Halo, 4.0 mm, Short, Model/Catalog Number: 40.070.NVG.H1; drills, burrs…Potential for burr breakage, occurring during bi-portal endoscopic spinal surgery (BESS). | 04/30/2026Open, Classified |
| Z-2561-2026 Class II | Surgify Halo, 4.0 mm, Short, Model/Catalog Number: 40.070.NVG.U1; drills, burrs…Potential for burr breakage, occurring during bi-portal endoscopic spinal surgery (BESS). | 04/30/2026Open, Classified |
| Z-2562-2026 Class II | Surgify Halo, 4.0 mm, Short, Model/Catalog Number: 40.125.NVG.H1; drills, burrs…Potential for burr breakage, occurring during bi-portal endoscopic spinal surgery (BESS). | 04/30/2026Open, Classified |
| Z-2563-2026 Class II | Surgify Halo, 40 mm, Long, Model/Catalog Number: 40.125.NVG.U1; drills, burrs, trephines…Potential for burr breakage, occurring during bi-portal endoscopic spinal surgery (BESS). | 04/30/2026Open, Classified |
| Z-2564-2026 Class II | Surgify Halo, 5.4 mm, Extendable, Model/Catalog Number: 54.000.SEE.H1; drills, burrs…Potential for burr breakage, occurring during bi-portal endoscopic spinal surgery (BESS). | 04/30/2026Open, Classified |
| Z-2565-2026 Class II | Surgify Halo, 5.4 mm, Extendable, Model/Catalog Number: 54.000.SEE.U1; drills, burrs…Potential for burr breakage, occurring during bi-portal endoscopic spinal surgery (BESS). | 04/30/2026Open, Classified |
| Z-2566-2026 Class II | Surgify Halo, 5.4 mm, Short, Model/Catalog Number: 54.070.NVG.H1; drills, burrs…Potential for burr breakage, occurring during bi-portal endoscopic spinal surgery (BESS). | 04/30/2026Open, Classified |
| Z-2567-2026 Class II | Surgify Halo, 5.4 mm, Short, Model/Catalog Number: 54.070.NVG.U1; drills, burrs…Potential for burr breakage, occurring during bi-portal endoscopic spinal surgery (BESS). | 04/30/2026Open, Classified |
| Z-2568-2026 Class II | Surgify Halo, 5.4 mm, Medium, Model/Catalog Number: 54.085.SHD.H1; burrs, trephines &…Potential for burr breakage, occurring during bi-portal endoscopic spinal surgery (BESS). | 04/30/2026Open, Classified |
| Z-2569-2026 Class II | Surgify Halo, 5.4 mm, Medium, Model/Catalog Number: 54.085.SHD.U1; drills, burrs…Potential for burr breakage, occurring during bi-portal endoscopic spinal surgery (BESS). | 04/30/2026Open, Classified |
| Z-2570-2026 Class II | Surgify Halo, 5.4 mm, Long, Model/Catalog Number: 54.125.NVG.H1; drills, burrs…Potential for burr breakage, occurring during bi-portal endoscopic spinal surgery (BESS). | 04/30/2026Open, Classified |
| Z-2571-2026 Class II | Surgify Halo, 5.4 mm, Long, Model/Catalog Number: 54.125.NVG.U1; drills, burrs…Potential for burr breakage, occurring during bi-portal endoscopic spinal surgery (BESS). | 04/30/2026Open, Classified |
| Z-2573-2026 Class II | Surgify Halo, 5.4 mm, Long, Model/Catalog Number: 54.140.SHD.U1; drills, burrs…Potential for burr breakage, occurring during bi-portal endoscopic spinal surgery (BESS). | 04/30/2026Open, Classified |