Recall Z-2597-2023
Stryker Corporation · initiated August 7, 2023
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Stryker ACM (Advanced Cement Mixing): 180-Gram Cement Cartridge with Breakaway Femoral Nozzle-Allows the user to mix bone cement with a constant high vacuum through the process Catalog Number: 0206-530-000
Codes / lots: GTIN: 04546540857880 Lot Numbers: 22332012 22336012 22333012 22355012 23052012
Reason for recall
Potential for the Break-Away Femoral Nozzle the applicator tip connecting the cement cartridge, to disassemble or break off during use, can lead to the potential for loss of function of the ACM system and bio-incompatibility due to components falling into surgical site
FDA-determined cause: Under Investigation by firm
Action taken
Stryker issued Urgent Medical Device Recall letter on 8/07/23. Letter states reason for recall, health risk and action to take: Actions needed 1. Immediately review your inventory to locate and quarantine any affected products at your facility. 2. Return the enclosed Business Reply Form (BRF), even if affected product is no longer in inventory, to the email Instruments.Recalls@stryker.com. 3. Upon receipt of the completed BRF, Stryker will provide a shipping label to return recalled product on-hand. 4. A replacement order will be shipped to customers upon receipt of the recalled product. 5. Maintain awareness of this communication internally and inform Stryker if any of the subject devices have been distributed to other organizations. If so, provide contact details so Stryker can inform the recipients accordingly. If you have any questions about this matter, contact Stryker Instruments by email at Instruments.Recalls@stryker.com.
Details
- Recall number
- Z-2597-2023
- Initiated
- August 7, 2023
- Posted by FDA
- September 19, 2023
- Product code
- LZN – Cement Obturator
- Quantity
- 224 units US; 718 units OUS
- Distribution
- Nationwide Foreign: Austria Belgium Canada France Germany Hong Kong Italy Malaysia Poland Singapore South Korea Spain Taiwan Thailand UK
- Recalling firm
- Stryker Corporation, Portage, MI
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-2596-2023 Class II | Stryker ACM (Advanced Cement Mixing) System: Break-Away Femoral Nozzle-, Allows the user…Potential for the Break-Away Femoral Nozzle the applicator tip connecting the cement cartridge, to… | 08/07/2023Open, Classified |
| Z-2598-2023 Class II | Stryker ACM (Advanced Cement Mixing) System: ACM Kit w/ Femoral Breakaway Nozzle-Allows…Potential for the Break-Away Femoral Nozzle the applicator tip connecting the cement cartridge, to… | 08/07/2023Open, Classified |
| Z-2599-2023 Class II | Stryker ACM (Advanced Cement Mixing) System: ACM Kit w/ Femoral Breakaway Nozzle & Solid…Potential for the Break-Away Femoral Nozzle the applicator tip connecting the cement cartridge, to… | 08/07/2023Open, Classified |
| Z-2600-2023 Class II | Stryker ACM (Advanced Cement Mixing) System: ACM Kit w/ Femoral Breakaway Nozzle & Prox.…Potential for the Break-Away Femoral Nozzle the applicator tip connecting the cement cartridge, to… | 08/07/2023Open, Classified |
| Z-2601-2023 Class II | Stryker ACM (Advanced Cement Mixing)System: ACM Kit w/ Femoral Breakaway Nozzle…Potential for the Break-Away Femoral Nozzle the applicator tip connecting the cement cartridge, to… | 08/07/2023Open, Classified |
| Z-2602-2023 Class II | Stryker ACM (Advanced Cement Mixing)System: ACM Kit w/ Femoral Breakaway Nozzle…Potential for the Break-Away Femoral Nozzle the applicator tip connecting the cement cartridge, to… | 08/07/2023Open, Classified |