Recall Z-2602-2026

Abiomed, Inc. · initiated May 22, 2026

Class I Open, Classified

Class I – dangerous or defective products that could cause serious health problems or death

Product

Abiomed Impella CP Set with SmartAssist (10th Generation) containing affected 14Fr Low Profile Introducer Kits 13 and 25 cm Japan. Product Code: 1000542.

Codes / lots: Impella Set Product Code: 1000542; GTIN: 813502013467; Serial numbers: N/A. Batch Number: 2041530. Introducer Product Code: 1000542; Batch Number: 2041530.

Reason for recall

Potential for thrombus formation during prolonged use of the introducer.

FDA-determined cause: Under Investigation by firm

Action taken

Abiomed notified consignees on about 05/22/2026, via letter titled "URGENT MEDICAL DEVICE RECALL (REMOVAL)." Consignees were made aware of the issue and that if thrombus formation were to occur, the user may encounter difficulty aspirating the sheath sidearm or valve and may observe thrombus in the syringe or sidearm post aspiration or around the valve post CP pump removal. They were instructed to review inventory for affected units and quarantine them, arrange for their return using the provided shipment label from Sedgwick, and complete and return the provided business response form. Consignees were also instructed to forward the notification to anyone in the facility that needs to be informed, if any of the subject products have been forwarded to another facility, contact that facility and provide them with the notification, and to post a copy of the notification in a visible area for awareness.

Details

Recall number
Z-2602-2026
Initiated
May 22, 2026
Posted by FDA
June 25, 2026
Product code
DYB – Introducer, Catheter
Quantity
7 units
Distribution
Worldwide - US Nationwide distribution in the states of FL, IL, MN, NJ, NY, VA and the country of Japan.
Recalling firm
Abiomed, Inc., Danvers, MA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2599-2026 Class IAbiomed 14 Fr x 13 cm Low Profile Introducer Kit for Impella CP. Impella Set Product…Potential for thrombus formation during prolonged use of the introducer. 05/22/2026Open, Classified
Z-2600-2026 Class IAbiomed 14 Fr x 25 cm Low Profile Introducer Kit for Impella CP. Impella Set Product…Potential for thrombus formation during prolonged use of the introducer. 05/22/2026Open, Classified
Z-2601-2026 Class IAbiomed Impella CP Set with SmartAssist (10th Generation) containing affected 14Fr Low…Potential for thrombus formation during prolonged use of the introducer. 05/22/2026Open, Classified