Recall Z-2603-2025

Philips North America Llc · initiated August 13, 2025

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Philips CT systems labeled as the following: 1. Ingenuity CT Family Brazil, Model Number: 728325.

Codes / lots: 1. Model Number: 728325; UDI-DI: N/A; Serial Numbers: 338014, 338028, 338019.

Reason for recall

The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.

FDA-determined cause: Process control

Action taken

Philips Medical notified consignees on about 08/11/2025 via letter. Consignees were instructed that they could continue using the affected systems accordance with the intended use and by following the recommendation: Observe the warning listed in Instructions for Use (IFU) and make sure that the patient is strapped securely to avoid dangling of the hands. Ensure that the patient is placed securely on the patient table and is not in danger of falling. Additionally, consignees were instructed to complete and return the provided response form, to circulate Urgent Medical Device Correction Letter to all users of this device so that they are aware of the issue and retain this letter with your system(s) until your system is inspected and corrected if necessary. Philips will contact consignees to schedule a time for a Philips Field Service Engineer (FSE) to visit each site and to inspect the affected system (FCO 72800832) and correct the system if necessary (FCO 72800833).

Details

Recall number
Z-2603-2025
Initiated
August 13, 2025
Posted by FDA
September 19, 2025
Product code
JAK – System, X-Ray, Tomography, Computed
Quantity
3 units
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Czech Republic, Denmark, Egypt, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lebanon, Mauritius, Mexico, Netherlands, Norway, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian…
Recalling firm
Philips North America Llc, Cambridge, MA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2595-2025 Class IIPhilips CT systems labeled as the following: 1. Brilliance CT 40 Channel, Model Number:…The patient support table (couch) may descend unexpectedly to the lowermost position due to a component… 08/13/2025Open, Classified
Z-2596-2025 Class IIPhilips CT systems labeled as the following: 1. Brilliance 16 Power, Model Number:…The patient support table (couch) may descend unexpectedly to the lowermost position due to a component… 08/13/2025Open, Classified
Z-2597-2025 Class IIPhilips CT systems labeled as the following: 1. Big Bore RT, Model Number: 728242. 2.…The patient support table (couch) may descend unexpectedly to the lowermost position due to a component… 08/13/2025Open, Classified
Z-2598-2025 Class IIPhilips CT systems labeled as the following: 1. Brilliance iCT, Model Number: 728306.The patient support table (couch) may descend unexpectedly to the lowermost position due to a component… 08/13/2025Open, Classified
Z-2599-2025 Class IIPhilips CT systems labeled as the following: 1. Brilliance iCT SP, Model Number: 728311.The patient support table (couch) may descend unexpectedly to the lowermost position due to a component… 08/13/2025Open, Classified
Z-2600-2025 Class IIPhilips CT systems labeled as the following: 1. Ingenuity CT 728326, Model Number:…The patient support table (couch) may descend unexpectedly to the lowermost position due to a component… 08/13/2025Open, Classified
Z-2601-2025 Class IIPhilips CT systems labeled as the following: 1. IQon Spectral CT, Model Number: 728332.The patient support table (couch) may descend unexpectedly to the lowermost position due to a component… 08/13/2025Open, Classified
Z-2602-2025 Class IIPhilips CT systems labeled as the following: 1. Ingenuity Core128 / Elite China, Model…The patient support table (couch) may descend unexpectedly to the lowermost position due to a component… 08/13/2025Open, Classified
Z-2604-2025 Class IIPhilips CT systems labeled as the following: 1. CT 6000, Model Number: 728307.The patient support table (couch) may descend unexpectedly to the lowermost position due to a component… 08/13/2025Open, Classified