Recall Z-2609-2023

Randox Laboratories Ltd. · initiated August 3, 2023

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Liquid Assayed Chemistry Control Premium Plus Level 3, Catalog Number LAE4215

Codes / lots: UDI-DI: 05055273208986; Lot Number: 1308UE

Reason for recall

there has been a transcription error in the Instructions for Use (IFU) for the Liquid Assayed Chemistry Control Premium Plus Level 3. For Caeruloplasmin, the one and two standard deviation values printed on the sheet are incorrect, which may lead to misreporting patient results. Target and range values are correct, and risk is mitigated; most clinical labs will use the target and range values in clinical practice.

FDA-determined cause: Under Investigation by firm

Action taken

A Medical Device Correction notification letter dated 8/3/23 was sent to customers. Action to be taken: " Review Quality Control data generated using the affected lot. " Discuss the contents of this notice with your Medical Director. " Complete and return the response form 12187-QA to technical.services@randox.com within five working days. Transmission of Field Safety Notice: Send a copy of the FSN to all affected customers and to those who need to be aware within your organisation. Please accept our apologies for any inconvenience caused. Thank you for your patience and understanding. If you have any questions or concerns please contact Randox Technical Services.

Details

Recall number
Z-2609-2023
Initiated
August 3, 2023
Posted by FDA
September 19, 2023
Product code
JJY – Multi-Analyte Controls, All Kinds (Assayed)
Quantity
66 kits
Distribution
US States: GA, OH, OR, PA, TX, VA
Recalling firm
Randox Laboratories Ltd., Crumlin, North

Official FDA recall record