Recall Z-2610-2010

Terumo Cardiovascular Corporation · initiated June 8, 2010

Terminated

Product

Sarns Cardioplegia Set Conducer Heat Exchanger Cplg/MP4 4:1 Bridge, Item# 6375

Codes / lots: Lot numbers: MD15, MD15A, MD22, MD29, MD29A, ME19, and ME26A

Reason for recall

Medical device vent port may be occluded and prevent delivery of priming fluid.

FDA-determined cause: Equipment maintenance

Action taken

Consignees were contacted on June 8, 2010, by phone, fax, email and priority mail to advise distributors and end-users of the potential occluded vent port defect and provide instructions for the replacement product recall. Consignees were mailed follow-up communications by priority overnight mail on June 9, 2010.

Details

Recall number
Z-2610-2010
Initiated
June 8, 2010
Posted by FDA
September 29, 2010
Terminated
February 29, 2012
Product code
DTR – Heat-Exchanger, Cardiopulmonary Bypass
Quantity
1682 kits
Distribution
The products were distributed nationwide.
Recalling firm
Terumo Cardiovascular Corporation, Elkton, MD

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2611-2010 Sarns Cardioplegia Set with Conducer Heat Exchanger Cplg/MP4 4:1, Item # 165720Medical device vent port may be occluded and prevent delivery of priming fluid. 06/08/2010Terminated
Z-2612-2010 Sarns Cardioplegia Set with w/ PVC Coil from MP4, Item # 15501Medical device vent port may be occluded and prevent delivery of priming fluid. 06/08/2010Terminated
Z-2613-2010 Sarns Cardioplegia Set CPLG BLD MP4 4:1 No recirculation, Item # 16010Medical device vent port may be occluded and prevent delivery of priming fluid. 06/08/2010Terminated
Z-2614-2010 Cardiovascular Procedure Kit X-Coated Pump and Table Pack, Item #71552Medical device vent port may be occluded and prevent delivery of priming fluid. 06/08/2010Terminated
Z-2615-2010 Cardiovascular Procedure Kit, (MP4 w/ Conducer Component Utilized), labeled in part…Medical device vent port may be occluded and prevent delivery of priming fluid. 06/08/2010Terminated
Z-2616-2010 Cardiovascular Procedure Kit Custom 4:1 Myocardial Set, (MP4 w/ Coil Component…Medical device vent port may be occluded and prevent delivery of priming fluid. 06/08/2010Terminated
Z-2617-2010 Cardiovascular Procedure Kit X-Coated H/L pack with Cardioplegia, Item # 70067-03Medical device vent port may be occluded and prevent delivery of priming fluid. 06/08/2010Terminated